Health O Meter 600KL-BT User Instructions page 13

Heavy duty eye level digital scale with integral digital height rod and built-in bluetooth wireless technology
Table of Contents

Advertisement

OPERATING INSTRUCTIONS (CONTINUED)
To Obtain BMI
1. Press the ON/OFF button to power on the scale.
2. Ask the patient to step on the scale platform and obtain the patient's weight. Press the
WEIGHT/HEIGHT/BMI button once. The scale will now display the position of the height rod.
While the patient remains on the platform, adjust the height rod so that the headpiece gently
rests on the patient's head. Press the WEIGHT/HEIGHT/BMI button again. The display will
show the patient's BMI.
3. If desired, press the SEND button to send the weight, height and BMI information to the
connected device.
4. Press the RESET button to revert back to weight display.
KG Only Procedure
Caution: Please make sure you carefully and precisely follow the directions below as you
will be in the PROGRAM MODE of your scale. Any modification outside of what is listed
below may greatly affect the functionality of your scale.
1. Press the ON/OFF button to power on the scale. As soon as you turn on the scale, press the
ZERO button once, the UNIT button twice, and then the ZERO button once again. The scale
will display "CAL".
2. Press the
once and "PROG" will show on the display.
3. Press the SEND button and OP1-1 will be displayed.
4. Press the
once and the display will read "OP2-0"for the KG only function, then press the
UNIT button to enable KG only mode. Your display should read "OP2-1".
5. Press the
(7 times) until display reads "DONE"
6. Press the SEND button, the scale will read "START", then read "0.0 Kg"
You have successfully locked your scale into KG Only.
KG Only (Deactivation) Procedure
1. Press the ON/OFF button to power on the scale. As soon as you turn on the scale, press the
ZERO button once, the UNIT button twice, and then the ZERO button once again. The scale
will display "CAL".
2. Press the
once and "PROG" will show on the display.
3. Press the SEND button and OP1-1 will be displayed.
once and the display will read "OP2-1"for the KG only function, then press the
4. Press the
UNIT button to disable KG only mode. Your display should read "OP2-0".
5. Press the
(7 times) until display reads "DONE"
6. Press the SEND button, the scale will read "START", then read "0.0"
You have successfully deactivated the KG Only feature on your scale.