Page 1
600KL-BT Heavy Duty Eye Level Digital Scale With Integral Digital Height Rod and Built-in ® BLUETOOTH Wireless Technology E112697 Patent: D667323 USER INSTRUCTIONS P/N UM600KL-BT Rev20180810...
® Thank you for your purchase of this Health o meter Professional scale. Please read this manual carefully and keep it for easy reference or training. TABLE OF CONTENTS CAUTIONS AND WARNINGS ..................3 DEFINITION OF SYMBOLS AND SPECIFICATIONS FOR THIS SCALE ....4 CONNECTIVITY / CERTIFICATIONS / DISPOSAL .............
INTENDED USE ® This Health o meter Professional platform scale is intended to be used in a professional medical environment by trained medical staff. This product was designed to weigh patients who are safely positioned and standing on the platform. Do not modify the product or use it for anything other than its intended purpose.
The weight can be displayed in pounds or in kilograms. The 600KL-BT features a swivel display which allows for viewing from either side of the scale plus wheels for easy mobility. The scale can be operated with a power adapter or by 6-C cell batteries (not included), depending on use.
Follow the national, regional or local regulations which ® apply to you for disposal. For scale assistance or telephone support, please call Health o meter Professional Customer Service (available on normal business days) at (+1) 800-815-6615 or (+1)
600KL-BT ASSEMBLY INSTRUCTIONS Before Assembly Each 600KL-BT Heavy Duty Eye Level Digital Scale with Integral Digital Height Rod is shipped disassembled in one carton. Carefully inspect the carton for shipping damage before unpacking. ® If damage is found, contact your shipper or a Health o meter Professional representative immediately at 1-800-815-6615.
Page 7
600KL-BT ASSEMBLY INSTRUCTIONS (continued) Prior to assembly, remove all brown cardboard in and around pillar, base, and display head. 1. Carefully remove the contents of the carton and place each part on a flat, level, and dry surface. Place the scale platform on its side and align the pillar perpendicular to the platform.
Page 8
600KL-BT ASSEMBLY INSTRUCTIONS (continued) 4. Plug the load cell cable connector into the load cell plug in the platform base. The base of the scale has been designed to provide wire management for the load cell cable. Route the load cell cable as shown.
Page 9
600KL-BT ASSEMBLY INSTRUCTIONS (continued) 7. Set the base/pillar assembly upright. Obtain the display head module. Plug the cable labeled “SCALE” into the SCALE jack in the display module. Plug the cable labeled “COM” into the COM jack in the display module.
SET UP Set Up 1. Remove the protective film from the display window on the display head. 2. Ensure the scale base is on a stable, level and flat surface and that the sides of the base are clear of any obstacles. 3.
OPERATING INSTRUCTIONS Using the Keypad FUNCTION ON/OFF Power scale ON and OFF. ZERO Zeroes the scale prior to weighing. Holds the value of the weighed object on the HOLD/RELEASE display until the button is pressed again to clear the value. UNITS Toggles between kilograms or pounds.
Page 12
OPERATING INSTRUCTIONS (CONTINUED) Height Measurement Instructions WARNING: Use caution when sliding height rod up and down. The rod should move smoothly. Do not move the rod in a jerking motion. Fold headpiece down when not in use. A patient must be at least 23-5/8” (60cm) tall for the height to be accurately measured. Note: 1.
Page 13
OPERATING INSTRUCTIONS (CONTINUED) To Obtain BMI 1. Press the ON/OFF button to power on the scale. 2. Ask the patient to step on the scale platform and obtain the patient’s weight. Press the WEIGHT/HEIGHT/BMI button once. The scale will now display the position of the height rod. While the patient remains on the platform, adjust the height rod so that the headpiece gently rests on the patient’s head.
BLUETOOTH® WIRELESS COMMUNICATION Note: The serial number on the Bluetooth® wireless dongle must match the BT device # on the label located on the back of the 600KL-BT scale display head. ® BLUETOOTH WIRELESS DONGLE SERIAL NUMBER AND BT DEVICE NUMBER MUST MATCH ®...
Hardware/Driver Installation (for Scale Communication to Windows® Download the CP210x USB to UART Bridge VCP Driver using one of the following methods: a. Go to http://www.silabs.com/products/mcu/pages/usbtouartbridgevcpdrivers.aspx b. Go to http://www.silabs.com/ and search for “CP2102 Driver Download”. Select the download for Windows® XP/Server 2003/Vista/7/8/8.1/10 (6.7) – Version 6.7 or the latest version available.
Page 16
Hardware/Driver Installation (for Scale Communication to Windows® c. Accept the license agreement and click “Next” d. Please wait while the driver installs.
Page 17
Hardware/Driver Installation (for Scale Communication to Windows® e. Driver installation is now complete. Click the “Finish” button to close the installation Wizard. Connect the USB Bluetooth® wireless dongle to any available USB 2.0 port on the Windows® PC.
Page 18
Windows® OS. Once the driver installation is complete, the Bluetooth® wireless dongle will automatically connect to the paired 600KL-BT scale and will be ready to transmit data. The driver installation is now ready to communicate with the following applications, Allscripts™...
(for Scale Communication to Welch Allyn CVSM) Note: To connect the scale to a Welch Allyn CVSM unit, Health o meter® Professional C- HOMWA-1 kit must be purchased separately. The HOMWA unit and the USB cable are not included with the 600KL-BT scale.
(for Scale Communication to Welch Allyn LXi) Note: To connect the 600KL-BT scale to a Welch Allyn LXi unit, a Health o meter® Professional L-HOMWA-1 kit must be purchased separately. The HOMWA unit and the RS232 cable (LXi cable) are not included in the 600KL-BT option box.
Page 21
Hardware Installation ® (for Scale Communication to Welch Allyn LXi) 3. Connect the RS232 cable male end from the HOMWA unit to “COM I” port on the back of the LXi. Back of LXi LXi “COM I” port 4. Power on the LXi unit and power on the scale to initialize Bluetooth® wireless communication.
Troubleshooting USB Bluetooth® Wireless Dongle USB Bluetooth® Wireless Dongle Fault Symptoms Problem Possible Cause Suggested Action Re-cycle power to the HOMWA unit by Power failure or USB disconnecting the USB cable No Power/USB connected to “TO CONNEX DEVICE” error failure or TO LXI and re-connecting USB Bluetooth®...
Caution: Disconnect the scale from the AC adapter power source prior to cleaning the unit. ® 1. Health o meter Professional recommends using one of the following solutions on a soft cloth or disposable wipe: ● mild soap and water solution ●...
CALIBRATION This scale is factory calibrated and does not require calibration prior to use. The calibration procedure may be performed using a known/certified weight in either pounds or kilograms. The default is in pounds. The following instructions assume the scale is set to pounds. If you prefer to calibrate in kilograms, please call Customer Service at 1-800-815-6615.
SCALE TROUBLESHOOTING Refer to the following instructions to check and correct any failure before contacting Customer Service at 1-800-815-6615. SYMPTOM POSSIBLE CAUSE CORRECTIVE ACTION Scale does not turn on 1. Dead battery 1. Replace batteries 2. Faulty electrical outlet 2. Use a different outlet 3.
HEIGHT ROD TROUBLESHOOTING SYMPTOM POSSIBLE CAUSE CORRECTIVE ACTION Questionable height 1. Improper connection 1. Check scale base 2. The display shows “___” connection, pillar 3. Height is out of connection, and display head connection – ensure tolerance cables are correctly connected 2.