Operation

Connecting to the Test Instrument

HVD series probes have been designed for use with Teledyne LeCroy oscilloscopes
equipped with the ProBus interface. When you attach the probe to the oscilloscope's input
connector, the instrument will:
Recognize the probe
Set the oscilloscope input termination to 1 MΩ
Activate the probe control functions in the oscilloscope user interface
Note: For accurate measurements, connect the probe to the oscilloscope and allow it to
warm up for at least 20 minutes.

Connecting to the Test Circuit

Two inputs are available at the probe tip to connect the probe to a circuit under test. For
accurate measurements, both the + and – inputs must be connected to the test circuit.
Positive voltages applied to the + input (red) relative to the – input (black) will deflect the
oscilloscope trace toward the top of the screen.
To maintain the probe's high performance, exercise care when connecting the probe.
Increasing the parasitic capacitance or inductance in the input paths may introduce a "ring"
or slow the rise time of fast signals. Input leads that form a large loop area will pick up any
radiated electromagnetic field that passes through the loop and may induce noise into the
probe inputs. Because this signal will appear as a differential mode signal, the probe's
common mode rejection will not remove it. This effect can be reduced by twisting the input
leads together to minimize the loop area.
WARNING. To avoid electric shock or fire, maintain the input leads in good
condition. The leads have a jacket wear indicator that shows through when the
jacket is excessively worn. If the white "WEAR" indicator on the input leads is visible,
cease use and contact Teledyne LeCroy service for repair or replacement.
High common mode rejection requires precise matching of the relative gain or attenuation
in the + and – input signal paths. Mismatches in additional parasitic capacitance,
inductance, delay, and a source impedance difference between the + and – signals will
lower the CMRR. Therefore, it is desirable to use the same length and type of wire and
connectors for both input connections. When possible, try to connect the inputs to points in
the circuit with approximately the same source impedance.
The probes may be stacked on one top of another during usage to conserve space. Since