HVD3000 Series High-Voltage Differential Probes
HVD3220 Specifications
For the current specifications, see the product datasheet at teledynelecroy.com.
Specifications are subject to change without notice.
G
S
UARANTEED
PECIFICATIONS
Bandwidth (probe only)
Risetime 10-90%
CMRR Test Limits, 23 C
E
C
LECTRICAL
HARACTERISTICS
Differential Voltage Range
(500X Attenuation)
Differential Voltage Range
(250X Attenuation)
Differential Voltage Range
(100X Attenuation)
Differential Voltage Range
(50X Attenuation)
Common Mode Voltage
(either input to ground)
Max. Input Voltage to Earth
(either input to ground)
Max. Safe Input Voltage*
(per IEC 61010-031)
Pollution Degree*
* See "IEC/EN 61010-031 Definitions" on p.40.
** CAT I per IEC/EN 61010-031/A1:2008. No Rated Measurement Category per IEC/EN 61010-031:2015.
V
S
ERTICAL
ENSITIVITY
500 mV/Div to 500 V/Div
20
HVD3220
400 MHz
875 ps
80 dB @ DC
60 dB @ 1 MHz
2000 V (DC + Peak AC)
from 35 V/div to 500 V/div, up to 1500 V offset
1000 V (DC + Peak AC)
from 14 V/div to 34.5 V/div, up to 750 V offset
400 V (DC + Peak AC)
from 7 V/div to 13.8 V/div, up to 300 V offset
200 V (DC + Peak AC)
from 500 mV/div to 6.9 V/div, up to 150 V offset
± 2000 V (DC + Peak AC)
± 2000 Vpk (nominal, either input to ground)
± 2000 Vpk 2000 V (DC + peak AC) CAT I **
1500 Vdc CAT III
1000 Vrms CAT III
2
EAR99 technology subject to restrictions on copyrights page.