User's Guide _______________________________________________________________________
Error Messages
74 ___________________________________________________________________ M211659EN-C
Table 13
Error Messages
Error Text
Error ID on
on Display
Serial Line
HTM10 01
23
HTM10 04
21
HTM10 03
4
HTM10 05
22
HTM10 06
42
GM10 01
79
GM10 03
81
GM10 04
82
GM10 10
28
GM10 11
88
GM10 12
89
Internal 1
1
Internal 2
2
Internal 3
3
Internal 4
43
Possible Cause
Remedy
Communication
Check that the HTM10
problem with
module sits firmly in place.
HTM10 module.
Remove and reconnect.
Problem with
-
humidity
measurement.
-
Internal problem
-
with HTM10
-
module.
Communication
Check that the GM10 module
problem with
sits firmly in place. Remove
GM10 module.
and reconnect.
Internal problem
-
with GM10
-
module.
Dirty or damaged
Replace module.
GM10 module.
Internal problem
-
with the
-
transmitter.
Check for missing or
®
damaged HUMICAP
sensor on the HTM10
module.
Check for condensation
®
on the HUMICAP
sensor. Wait for the
sensor to dry out.
Restart the transmitter.
Replace the module if
unable to remove the
problem.
Restart the transmitter.
Replace the module if
unable to remove the
problem.
Restart the transmitter.
Restore the factory
settings using service
port or DIP switches if
reset does not help.