Figure 29 Foam Pipe Inside Gmw90 - Vaisala CARBOCAP GMW90 Series User Manual

Carbon dioxide, temperature, and humidity transmitters
Hide thumbs Also See for CARBOCAP GMW90 Series:
Table of Contents

Advertisement

Chapter 5 _______________________________________________________________ Maintenance
VAISALA ________________________________________________________________________ 71
4.
A short piece of foam pipe is included with the GM10 spare part.
Place it on top of the calibration gas inlet inside the transmitter, as
shown in the figure below. If the foam pipe is already in place and
appears undamaged, you do not have to replace it.
Figure 29
5.
Take the new GM10 module and place it in the hole for the module
so that the plastic clip meets the component board. Then lower the
connector end of the module.
6.
Push down on the module to secure the connector.
7.
Reconnect the transmitter to the mounting base.
8.
Verify that there are no errors when the transmitter starts up. If
there are, see section Error Messages on page 74.
Foam Pipe Inside GMW90
1602-55