Repair Maintenance; Replacing The Gm10 Module; Figure 28 Disconnecting The Gm10 Module - Vaisala CARBOCAP GMW90 Series User Manual

Carbon dioxide, temperature, and humidity transmitters
Hide thumbs Also See for CARBOCAP GMW90 Series:
Table of Contents

Advertisement

User's Guide _______________________________________________________________________

Repair Maintenance

NOTE
70 ___________________________________________________________________ M211659EN-C
If you cannot restore the measurement accuracy of the transmitter by
calibration and adjustment, you can replace the measurement modules
inside the transmitter. Measurement modules are the small component
boards that are connected to the main transmitter component board. See
Figure 4 on page 17.
User calibration settings (adjustment by trimmers or service port) are
stored in the module. If you replace the module, you do not need to undo
the previously applied correction.

Replacing the GM10 Module

To replace the module:
1.
Disconnect the transmitter body from the mounting base.
2.
Locate the GM10 module on the transmitter. The module is a
separate component board with a golden cuvette that contains the
®
CARBOCAP
transmitter board with a connector, and held in place by a plastic
clip on the other side. See Figure 4 on page 17.
3.
Disconnect the module by carefully lifting the module from the
connector side. See Figure 28 below.
Figure 28
sensor. The module is connected to the main
Disconnecting the GM10 Module
1311-137