Adjustment Using Display and Trimmers

The trimmers only turn 135 degrees each way, less than half a rotation.
Do not force the trimmer past the stopping point. If you wish to apply a
greater correction than allowed by the trimmer in a single adjustment,
simply adjust the transmitter again. Corrections applied using the
trimmers are cumulative.
User calibration settings (adjustment by trimmers or service port) are
stored in the measurement modules. If you replace a module, there is no
need to undo previous adjustments.
1.
To enter the adjustment screen, open the slide and rotate the CO
T, or RH trimmer slightly during normal measurement. If the
trimmer is not centered, you see the trimmer centering screen first.
Simply turn the trimmer to the center and wait for the progress bar
to complete.
1111-075
Figure 26
2.
In the adjustment screen, turn the trimmer to set the desired
correction. To commit the change, stop turning the trimmer and
wait.
1310-092
Figure 27
3.
If you have adjusted the CO
for the reading to stabilize.
4.
Verify the adjusted measurement from measurement screen.
Trimmer Centering Screen
Adjustment Screen
measurement, wait for three minutes