Notes For T Adjustment; Notes For Rh Adjustment; Figure 25 Effect Of Pressure On Co Reading - Vaisala CARBOCAP GMW90 Series User Manual

Carbon dioxide, temperature, and humidity transmitters
Hide thumbs Also See for CARBOCAP GMW90 Series:
Table of Contents

Advertisement

User's Guide _______________________________________________________________________
66 ___________________________________________________________________ M211659EN-C
When adjusting the transmitter, you can also correct your reference
concentration (for example, the value on the gas cylinder label) according
to the graph below.
Figure 25
Effect of Pressure on CO

Notes for T Adjustment

Temperature adjustment is always a simple 1-point offset correction.

Notes for RH Adjustment

1-point humidity adjustment affects both offset and gain, depending on
the adjustment condition. In a dry condition (for example, 11 %RH),
offset is adjusted more than gain.
The 1-point humidity adjustment requires that the target condition is at
least 50% of the currently measured value. This prevents the user from
making excessive corrections that are very unlikely to be needed.
Reading
2
1402-025

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents