Chapter 4 _________________________________________________________________ Operation
NOTE
VAISALA ________________________________________________________________________ 55
Clear User Adjustment of CO
CCO2 [RESET]<cr>
Example:
>pass 9000
>cco2 reset
OK

Adjust Humidity Measurement

Use the CRH command to perform a humidity adjustment of the relative
humidity (RH) measurement. You can do a 1-point or a 2-point
adjustment, or clear the adjustment information from the HTM10
module. Note that the factory adjustment will remain intact when user
adjustment is cleared.
Before using the CRH command, read section Notes for RH Adjustment
on page 66.

Show Current RH Adjustment

CRH<cr>
Example (showing default offset and gain):
>pass 9000
>crh
RH Gain
RH Offset
Measurement
2
: 1.000
: 0.000