User's Guide _______________________________________________________________________
44 ___________________________________________________________________ M211659EN-C
Example (show current configuration):
>pass 9000
>amode
Aout 1 range (mA)
Aout 2 range (mA)
Aout 3 range (mA)
Example (set channel 1 to 0 ... 20 mA output, with error level at 21 mA):
>amode 1 0 20 21
Aout 1 range (mA)
>

Set Analog Output Scaling

Use the ASEL command to select the output parameter and scaling for
analog output channels.
ASEL [channel parameter lo_value hi_value]<cr>
where
channel
= Analog output channel, 1 ... 3.
parameter = Parameter that is output on the channel.
Available parameters are:
CO2
RH
T
Tdf
Td
Tw
h
x
a
dTd
lo_value
= Low limit of the scaling, in the units of the selected
parameter.
hi_value
= High limit of the scaling in the units of the selected
parameter.
Example (set channel 1 to output dewpoint, in the range -10 ... 20 °C):
>pass 9000
>asel 1 TD -10 20
Aout 1 quantity
: 4.00 ... 20.00 (error: 3.60)
: 4.00 ... 20.00 (error: 3.60)
: 4.00 ... 20.00 (error: 3.60)
: 0.00 ... 20.00 (error: 21.00)
carbon dioxide
relative humidity
temperature
dew/frostpoint temperature
dewpoint temperature
wetbulb temperature
enthalpy
mixing ratio
absolute humidity
dew/frostpoint depression
: Td (-10.00 ... 20.00 'C)