Modbus Registers; Measurement Data Registers; Status Registers - Vaisala CARBOCAP GMW80 Series Quick Manual

Carbon dioxide and temperature transmitter
Hide thumbs Also See for CARBOCAP GMW80 Series:
Table of Contents

Advertisement

Value
8002
... FFFF
hex
hex
8000
hex
The 16-bit integer value for measured temperature in
the necessary decimals.

Modbus registers

CAUTION!
addresses in actual Modbus messages (Modbus Protocol Data Unit (PDU)) are in
hexadecimal and start from zero. Subtract 1 from the register number presented
in this manual to get the address used in the Modbus message (for example,
register number 1 corresponds to address 0
Accessing unavailable (temporarily missing) measurement data does not generate a Modbus
exception. "Unavailable" value (a quiet NaN for floating point data or 8000
is returned instead. An exception is generated only for any access outside the applicable
register ranges.

Measurement data registers

Table 4 Modbus measurement data registers (read-only)
Register
Address
number
1
0000
hex
5
0004
hex
257
0100
hex
259
0102
hex

Status registers

Table 5 Modbus status registers (read-only)
Register
Address
number
513
0200
hex
514
0201
hex
10
Registers are numbered in decimal, starting from 1. Register
Register description
Measured CO
value
2
Measured temperature
Measured CO
value
2
Measured temperature (T * 10) 16-bit integer
Register description
Device status
Error code
Description
Value in range -32766 ... -1 (2's complement)
Value is not available
Table 4 (page 10)
in the actual Modbus message).
hex
Data format
32-bit float
32-bit float
16-bit integer
Data
Notes
format
16-bit
0000
: Status OK; if other values
hex
integer
are returned, see
(page
8).
32-bit
Used by Vaisala Technical Support
integer
when solving device problems
is scaled to include
for integer data)
hex
Unit
ppm
°C
ppm
°C
Troubleshooting