Vaisala CARBOCAP GMW80 Series Quick Manual

Vaisala CARBOCAP GMW80 Series Quick Manual

Carbon dioxide and temperature transmitter
Hide thumbs Also See for CARBOCAP GMW80 Series:

Advertisement

Quick Links

M212445EN-A
Quick Guide
â
Vaisala CARBOCAP
Carbon Dioxide
and Temperature Transmitter
GMW87M

Advertisement

Table of Contents
loading

Summary of Contents for Vaisala CARBOCAP GMW80 Series

  • Page 1 M212445EN-A Quick Guide â Vaisala CARBOCAP Carbon Dioxide and Temperature Transmitter GMW87M...
  • Page 2 Vaisala Oyj Vanha Nurmijärventie 21, FI-01670 Vantaa, Finland P.O. Box 26, FI-00421 Helsinki, Finland +358 9 8949 1 Visit our Internet pages at www.vaisala.com. © Vaisala 2020 No part of this document may be information contained in this document. reproduced, published or publicly...
  • Page 3: Table Of Contents

    Table of contents Product overview....................3 Transmitter parts..........................3 Safety.............................. 4 Regulatory compliances......................4 Installation......................5 Wiring............................. 6 Modbus communication...................7 Troubleshooting........................... 8 Modbus reference....................9 Function codes..........................9 Data encoding..........................9 32-bit floating point format....................9 16-bit integer format......................9 Modbus registers........................10 Measurement data registers................... 10 Status registers........................
  • Page 4 M212445EN-A...
  • Page 5: Product Overview

    Product overview Vaisala CARBOCAPâ Carbon Dioxide and Temperature Transmitter GMW87M is a wall-mount transmitter designed to fulfill the needs for CO measurements in demanding conditions (for example, dusty or humid installation locations). GMW87M can be accessed with the Modbus RTU protocol over an RS-485 interface. The communication settings for the protocol are entered with DIP switches on the GMW87M component board.
  • Page 6: Safety

    Electrostatic Discharge (ESD) can cause immediate or latent damage to electronic circuits. Avoid touching exposed component contacts during installation and maintenance. Regulatory compliances Vaisala GMW80 series transmitters are in conformity with the provisions of the following EU directives: • RoHS Directive • EMC Directive The conformity is declared with using the following standards: •...
  • Page 7: Installation

    Installation The GMW87 transmitter is designed to be installed on a mounting box or directly to a wall surface. For transmitter dimensions, see Figure 3 (page Figure 2 Mounting GMW87M Cable gland for leading the input/output cable (Ø 4 ... 8 mm) inside the enclosure Alternative lead-through for wiring through the back (break seal and transfer cable gland to the threads on this lead-through) Wiring instructions label...
  • Page 8: Wiring

    6. Push down gently on the GM10 module. The module should be straight and firmly connected. 7. Tighten the cable gland and close the transmitter cover (4 screws). CAUTION! Mounting bases are tailored to the individual transmitter at the factory. If you attach several mounting bases before connecting the transmitters, store the transmitters in their original shipping boxes, so that you can match the serial number on the mounting base to the serial number on the shipping box.
  • Page 9: Modbus Communication

    Modbus communication The Modbus variant used in GMW87M is Modbus RTU. For a list of the Modbus registers available in GMW87M, see Modbus registers (page 10). You can use up to 8 transmitters on the same RS-485 line. You must configure each transmitter on the line to have a different Modbus address.
  • Page 10: Troubleshooting

    • Check that the GM10 module is properly attached • Contact Vaisala for a replacement GM10 module If you cannot locate the source of the error and return the transmitter to operational state, contact Vaisala technical support. See Technical support (page 15).
  • Page 11: Modbus Reference

    Modbus reference Function codes Table 2 Modbus function codes Function code (decimal) Function code (hexadecimal) Name Read Holding Registers 43/14 / 0E Read Device Identification Data encoding In the data registers, the numeric values are available in one or two formats with separate register addresses: 32-bit IEEE floating point format and/or 16-bit signed integer format.
  • Page 12: Modbus Registers

    Register Address Register description Data Notes number format 0200 Device status 16-bit 0000 : Status OK; if other values integer are returned, see Troubleshooting (page 0201 Error code 32-bit Used by Vaisala Technical Support integer when solving device problems M212445EN-A...
  • Page 13: Test Value Registers

    Table 7 Device identification objects Object ID Object ID Object name Example contents (hexadecimal) VendorName “Vaisala” ProductCode "GMW80" MajorMinorVersion Software version (for example, "1.4.0") VendorUrl “https://www.vaisala.com/” ProductName "Vaisala CARBOCAP(R) Carbon Dioxide Transmitter GMW80" Transmitter serial number, e.g. SerialNumber "J1140501" Vaisala-specific device information object.
  • Page 14: Modbus Communication Example

    Modbus communication example Reading CO concentration value The device address used in the following example is 240 (F0 ). The values returned by the device differ depending on the ambient conditions. Your device might not return exactly the same values. Request Response Bytes on the line...
  • Page 15: Technical Data

    ±(15 ppm + 2 % of reading) over five years Warm-up time 10 min for full specification Response time (63 %) 7 min Carbon dioxide sensor Vaisala CARBOCAPâ GM10 Product lifetime > 15 years Table 9 Operating environment Property Specification Operating temperature 0 …...
  • Page 16 Property Specification Digital output (RS-485) Isolated, supports Modbus RTU protocol • Modbus RTU address range: • 1 … 247 (up to 255 possible, non-standard) • Bit rates: • 4800, 9600, 19200, 38400, 57600, 76800, 115200 • Parity: • None or Even •...
  • Page 17: Warranty

    Please see the applicable supply contract or Conditions of Sale for details of the warranty for each product. Technical support Contact Vaisala technical support at helpdesk@vaisala.com. Provide at least the following supporting information as applicable: • Product name, model, and serial number •...