8
Meter Maintenance and Troubleshooting
Meter Maintenance
The meter automatically tests its own
systems every time you turn it on and lets
you know if something is wrong. See the
Error Messages section in this chapter.
If you have problems with the meter, drop
the meter, or think the results are not
accurate, perform a control test with an
unexpired test strip and control solution. If
the control result is not within the
acceptable range, contact the Accu‑Chek
Customer Care Service Center at
1‑800‑858‑8072.
WARNING
Keep new and used batteries away
from children. See the warning in the
Important Safety Information section at
the beginning of this manual for
additional information.

Changing the Batteries

1
Open the battery door
on the back of the
meter by pushing the
tab in the direction of
the arrow and pulling
the door up.
2
Remove the old
batteries.
3
back in place and snap it closed.
NOTE
• Always have a spare set of batteries.
• Battery life may vary due to factors
such as temperature and battery
manufacturer.
• If the E‑7 or E‑9 error code still
appears on the display after you have
changed the batteries, remove the
batteries again, press
seconds, then reinsert the batteries.
• The meter uses two 3‑volt lithium
batteries, coin cell type CR2032. This
type of battery can be found in many
stores.
• Always replace both batteries at the
same time and with the same brand.
• Meter data is not lost when you
replace the batteries.

Error Messages

WARNING
• Never make therapy decisions based
on an error message.
• If you have any concerns or see any
other error message, contact the
Accu‑Chek Customer Care Service
Center at 1‑800‑858‑8072.
31
Slide the new batteries
under the black tabs
with the (+) side facing
up.
Put the battery door
for at least 2