CALIBRATION

This scale has been factory calibrated, and does not require calibration prior to
use. If required, the scale can be calibrated. Calibration of your scale must be
performed using kilograms (KG). 20 kilograms are required for calibration, applied in 5
kg increments.
Note: Only weights certified and traceable to national standards should be used
for calibration procedures.
With the scale off, press and hold the LB/KG button (522KL) or the
blank blue button (522KG) and ZERO/TARE buttons at the same
time, then press the ON/OFF button to turn on the scale. "CAL" will
briefly appear on the display, followed by four numbers indicating that
you can begin calibration. Release all three buttons.
Press the ZERO/TARE button. "C-5" will flash on the display,
followed by "0"or "1".
Place 5 kg on the scale. Four numbers will appear. When calibration
numbers stabilize, press the ZERO/TARE button. "C-10" will appear
on the display, followed by "0" or "1".
Place an additional 5 kg on the scale, for a total of 10 kg. Four
numbers will appear. When calibration numbers stabilize, press the
ZERO/TARE button. "C-15" will appear on the display, followed by "0"
or "1".
Place an additional 5 kg on the scale, for a total of 15 kg. Four
numbers will appear. When calibration numbers stabilize, press the
ZERO/TARE button. "C-20" will appear on the display, followed by "0"
or "1".
Place an additional 5 kg on the scale, for a total of 20 kg. Four
numbers will appear. When calibration numbers stabilize, press the
ZERO/TARE button. "END" will appear on the display, followed by
"OFF". The scale will automatically shut off.
P/N UM522KL_522KG Rev20190412
ACTION
13
DISPLAY
SHOWS
"CAL" followed by
"XXXX"
"C-5" then "0" or
"1".
"XXXX" followed
by "C-10" then "0"
or "1".
"XXXX" followed
by "C-15" then "0"
or "1".
"XXXX" followed
by "C-20" then "0"
or "1".
"XXXX" followed
by "END" then
"OFF"