Teledyne Lecroy AP031 Operator's Manual

Teledyne Lecroy AP031 Operator's Manual

Differential probe

Advertisement

Quick Links

Operator's
Manual
AP031
Differential Probe
1.800.561.8187
information@itm.com
www.
.com

Advertisement

Table of Contents
loading

Summary of Contents for Teledyne Lecroy AP031

  • Page 1 Operator’s Manual AP031 Differential Probe 1.800.561.8187 information@itm.com www. .com...
  • Page 2: Warranty

    Spare parts, replacement parts and repairs are warranted for 90 days. In exercising its warranty, Teledyne Lecroy, at its option, will either repair or replace any assembly returned within its warranty period to the Customer Service Department or an authorized service center. However, this will be done only if the product is determined by Teledyne Lecroy’s examination to be defective due to workmanship or...
  • Page 3: Safety Instructions

    AP031 Probe Safety Instructions This section contains instructions that must be observed to keep this oscilloscope accessory operating in a correct and safe condition. You are required to follow generally accepted safety procedures in addition to the precautions specified in this section. The overall safety of any system incorporating this accessory is the responsibility of the assembler of the system.
  • Page 4 Operator’s Manual Precautions To avoid personal injury or damage to property, review and comply with the following safety precautions. Use product only as specified. The probe is designed to make differential voltage measurements. It is not to be used to insulate the circuit under test from the measuring instrument.
  • Page 5 If any part is damaged, cease operation immediately and sequester the probe from inadvertent use. Introduction The AP031 is a fully differential active probe designed for applications where electrical signals must be measured relative to a floating voltage different to the oscilloscope ground potential.
  • Page 6 Operator’s Manual voltage is calculated within the probe, with only the resultant difference signal being passed to the oscilloscope, a large dynamic range can be achieved with excellent rejection of common mode signals. Figure 1 923064-00 Rev A 1.800.561.8187 information@itm.com www.
  • Page 7: Specifications

    AP031 Probe Specifications Electrical Characteristics Bandwidth 25 MHz Rise Time 14 ns Attenuation 1:10 / 1:100 Atten. Accuracy ±2 % Input Resistance 4 MΩ Input Capacitance 5.5 pF each side to ground Input Configuration Differential Input Voltage Max. Differential 1:100 Range ±700 V (DC + AC pk) and 700 V r.m.s.
  • Page 8: Electrical Ratings

    Operator’s Manual Electrical Ratings Maximum Input Voltage 1000 V rms (1400 V DC + peak AC) Installation (Overvoltage) Category Pollution Degree Measurement Category III (CAT III) is applicable for measurements performed in the building installation. Pollution Degree 2 is applicable to an operating environment where normally only dry non-conductive pollution occurs.
  • Page 9 AP031 Probe Voltage vs. Frequency Derating NOTE: The voltage vs frequency derating curve provides the maximum voltage that can be applied to the probe inputs without risking damage to the probe or injury to the user from possible burns at higher frequencies.
  • Page 10: Making Measurements

    To ensure the oscilloscope correctly interprets the vertical waveform scale be sure to adjust the probe attenuation setting using the ‘Probe Atten’ menu controls. The example shown, x100 probe attenuation, would achieve correct vertical scaling AP031 operating with a maximum range of ± 700 V.
  • Page 11: Maintenance

    The AP031 series probes utilize fine-pitch surface mount devices. It is, therefore, impractical to attempt repair in the field. Defective probes must be returned to a Teledyne LeCroy service facility for diagnosis and exchange. A defective probe under warranty will be replaced with a factory refurbished probe.
  • Page 12: Emc Compliance

    Operator’s Manual Certifications This section certifies the AP031 probe’s Electromagnetic Compatibility (EMC), Safety and Environmental compliances. EMC Compliance EC Declaration of Conformity - EMC The probe meets intent of EC Directive 2004/108/EC for Electromagnetic Compatibility. Compliance was demonstrated to the following specifications as listed in the Official Journal...
  • Page 13: Australia & New Zealand Declaration Of Conformity-Emc

    AP031 Probe European Contact: Teledyne LeCroy Europe GmbH Waldhofer Str 104 D-69123 Heidelberg Germany Tel: (49) 6221 82700 Australia & New Zealand Declaration of Conformity—EMC The Probe complies with the EMC provision of the Radio Communications Act per the following standards, in accordance with requirements imposed by Australian...
  • Page 14: Safety Compliance

    Operator’s Manual Safety Compliance EC Declaration of Conformity – Low Voltage The probe meets intent of EC Directive 2006/95/EC for Product Safety. Compliance was demonstrated to the following specifications as listed in the Official Journal of the European Communities: EN 61010-1:2010 Safety requirements for electrical equipment for measurement, control, and laboratory use –...
  • Page 15: Environmental Compliance

    Many countries prohibit the disposal of waste electronic equipment in standard waste receptacles. For more information about proper disposal and recycling of your Teledyne LeCroy product, Restriction of Hazardous Substances (RoHS) This probe has been classified as Industrial Monitoring and Control Equipment, and is outside the scope of the 2011/65/EU RoHS Directive until 22 July 2017 (per Article 4, Paragraph 3).

Table of Contents