CALIBRATION

This scale has been factory calibrated, and does not require calibration prior to use.
There are no requirements for calibration in the field; users should adhere to their
institution's calibration policies.
Calibration of your scale can be performed using calibrated kilogram (kg) or pound (lb)
weight standards. Minimum calibration is 200 lb / 100 kg. Maximum calibration weight is
1000 lb / 454 kg. The required calibration standard will be the Unit of Measure (UOM)
for which the scale is configured. KG models can only be calibrated in kilograms. If
®
Everlock
is engaged the scale can only be calibrated in the locked unit of measure.
Note: Only weights certified and traceable to national standards should be used for
calibration.
Warning: Always ensure that the scale is returned to the correct UOM after scale
calibration.
Help
1. Press
to access the Help menu options.
2. While in the Help menu, press
Press
to advance to the next step.
3. The display will prompt for entry of the password. Press
the following keys in this sequence to proceed with the
calibration procedure.
4. The display will show a warning. If prepared and
qualified to perform the calibration procedure,
press
to advance to the next step.
5. Follow the screen prompt and ensure the scale is on a
level surface with no weight on the platform.
Press
to advance to the next step.
6. The display will show "Zeroing Scale" as it calculates the
zero value. Do not touch the scale during the zeroing
process.
P/N UM1100KL_1100KG_REV20181128
to select "Calibrate".
22