Teledyne Lecroy CP031 Instruction Manual

Current probe
Hide thumbs Also See for CP031:

Advertisement

Quick Links

Instruction Manual
CP031 Current Probe
1.800.561.8187
information@itm.com
www.
.com

Advertisement

Table of Contents
loading

Summary of Contents for Teledyne Lecroy CP031

  • Page 1 Instruction Manual CP031 Current Probe 1.800.561.8187 information@itm.com www. .com...
  • Page 2 CP031 Current Probe Instruction Manual August 2013 1.800.561.8187 information@itm.com www. .com...
  • Page 3 Spare parts, replacement parts and repairs are warranted for 90 days. In exercising its warranty, Teledyne Lecroy, at its option, will either repair or replace any assembly returned within its warranty period to the Customer Service Department or an authorized service center. However, this will be done only if the product is determined by Teledyne LeCroy’s examination to be defective due to workmanship or...
  • Page 4: Table Of Contents

    Precautions ................................4 Connecting the Probe to the Test Instrument ......................4 Connecting the Probe to the Test Circuit ........................4 Operation with a Teledyne LeCroy Oscilloscope ...................... 5 Bandwidth Limit ................................. 5 Auto Zero .................................. 5 Degauss Probe ................................. 5 Care and Maintenance ..............................6...
  • Page 5 Safety Compliance ..............................18 EC Declaration of Conformity – Low Voltage ...................... 18 Environmental Compliance ............................. 18 End-Of-Life Handling ............................18 Restriction of Hazardous Substances (RoHS) ....................18 Contact Teledyne LeCroy ............................. 19 Appendix A ................................... 20 923362-00 Rev A 1.800.561.8187 information@itm.com www.
  • Page 6: Safety Instructions

    Instruction Manual Safety Instructions This section contains instructions that must be observed to keep this oscilloscope accessory operating in a correct and safe condition. You are required to follow generally accepted safety procedures in addition to the precautions specified in this section. The overall safety of any system incorporating this accessory is the responsibility of the assembler of the system.
  • Page 7: Operating Environment

     To avoid short circuits and accidents that could result in injury or death, use the CP031 current probe only with conductors carrying 300 V or less. ...
  • Page 8: Overview

    The probe can be used with a WaveSurfer, WaveRunner 6000, or WavePro 7000 series oscilloscope with firmware 4.3.0.0 or higher. With the ProBus interface, the CP031 becomes an integral part of the oscilloscope. The bandwidth limit, auto zero and degauss functions are all controlled from the oscilloscope’s graphical user interface.
  • Page 9: Operation

    To use the CP031 with a WaveMaster scope, the AP-1M impedance adapter is required. Connecting the Probe to the Test Circuit The CP031 has been designed with a movable split core, eliminating the need to break the conductor for the core to slip around the conductor.
  • Page 10: Operation With A Teledyne Lecroy Oscilloscope

    Turning the VOLTS/DIV knob will control the oscilloscope’s scale factor to give full dynamic range from 20 mA/div to 50 A/div. The CP031 probe dialog (a tab behind the channel setup dialog) allows for the selection of the probe’s coupling (DC, Grounded or AC), AUTO ZERO, DEGAUSS PROBE and Probe BWL functions, limiting the system bandwidth to 20 MHz or maximum bandwidth (BWL Off).
  • Page 11: Care And Maintenance

    (A Performance Verification / Adjustment procedure is included in this manual.) Service Strategy Defective probes must be returned to a Teledyne Lecroy service facility for diagnosis and exchange. A defective probe under warranty will be replaced with a factory refurbished probe. A probe that is not under warranty can be exchanged for a factory refurbished probe.
  • Page 12: Returning A Defective Probe

    5. Package the probe case in a cardboard shipping box with adequate padding to avoid damage in transit. 6. Mark the outside of the box with the shipping address given to you by the Teledyne Lecroy representative; be sure to add the following: •...
  • Page 13: Performance Verification

    CP031 Current Probe Performance Verification This procedure can be used to verify the warranted characteristics of the CP031 Current Probe. The recommended calibration interval for the model CP031 Current Probe is one year. The complete performance verification procedure should be performed as the first step of annual calibration.
  • Page 14: Preliminary Procedure

    1. Connect the CP031 to the channel 1 input of the oscilloscope and completely close the probe slider. 2. Turn the oscilloscope on and allow at least 30 minutes warmup time for the CP031 and test equipment before performing the Verification Procedure.
  • Page 15: Check Lf Accuracy

    6. With the CP031 removed from any signal and the slider returned to the LOCKED position, degauss the probe by pressing the DEGAUSS button (located in the CP031 dialog), then OK. 7. Open the CP031 slider and position the probe input around the 100 Turn loop. Close and LOCK the slider.
  • Page 16: Adjustment Procedure

    Instruction Manual Adjustment Procedure This procedure can be used to adjust the warranted characteristics of the CP031 Current Probe. This procedure should be used if a parameter measured in the Performance Verification Procedure is outside of the specification limits. Adjustment should only be performed by qualified personnel.
  • Page 17: Preliminary Procedure

    2. Connect the CP031 compensation board to the channel 1 input of the oscilloscope through the ProBUS extension cable, and completely close the probe slider. 3. Turn the oscilloscope on and allow at least 30 minutes warmup time for the CP031 and test equipment before performing the Verification Procedure.
  • Page 18 6. With the CP031 removed from any signal and the slider returned to the LOCKED position, degauss the probe by touching DEGAUSS (on the CP031 dialog), then OK. 7. Open the CP031 slider and position the probe input around the 100 Turn loop. Close and LOCK the slider.
  • Page 19: Specifications

    CP031 Current Probe Specifications Nominal Characteristics Nominal characteristics describe parameters and attributes which are guaranteed by design, but do not have associated tolerances. Maximum Continuous Input Current 30 A (Refer to Figure 3) Maximum Peak Current 50 A , noncontinuous...
  • Page 20: Environmental Characteristics

    Instruction Manual Environmental Characteristics Operating Temperature Humidity 0 to 40 °C (32 °F to 104 °F) ≤ 80% relative humidity (non-condensing) Storage Temperature Humidity –10 °C to 50 °C (14 °F to 122 °F) ≤ 80% relative humidity (non-condensing) Usage Indoor Altitude up to 2000 m (6562 feet)
  • Page 21: Graphs

    CP031 Current Probe Graphs Current (Arms) 1.00E+02 1.00E+03 1.00E+04 1.00E+05 1.00E+06 1.00E+07 1.00E+08 Frequency (Hz) Figure 3. Maximum input current vs. frequency. Figure 4. Insertion impedance vs. frequency. 923362-00 Rev A 1.800.561.8187 information@itm.com www. .com...
  • Page 22: Certifications

    3 Meets Performance Criteria “B” limits of the respective standard: during the disturbance, product undergoes a temporary degradation or loss of function or performance which is self-recoverable. UROPEAN ONTACT Teledyne LeCroy Europe GmbH Waldhofer Str 104 D-69123 Heidelberg Germany Tel: (49) 6221 82700 Australia &...
  • Page 23: Safety Compliance

    CP031 Current Probe CISPR 11:2003 Radiated and Conducted Emissions, Group 1, Class A, in accordance with EN61326-1:2006 and EN61326-2-1:2006. USTRALIA EALAND ONTACTS Vicom Australia Ltd. Vicom New Zealand Ltd. 1064 Centre Road 60 Grafton Road Oakleigh, South Victoria 3167 Auckland...
  • Page 24: Contact Teledyne Lecroy

    Performance Verification Test Record Photocopy this page and record the results of measurements made during the performance verification of the CP031 Current Probe on the copy. File the completed record as required by applicable internal quality procedures. The sections in the test record correspond to the parameters tested in the performance verification procedure.