A.4.2 Status Registers
Table 52 Modbus Status Registers (Read-Only)
Register
Address
Number
(Hexadecimal)
(Decimal)
513
0200
hex
516
0203
hex
518
0205
hex
Table 53 Error Codes on Modbus Interface
Error Code
Register Value
1
Temperature measurement error. [44]
2
Humidity measurement error. [45]
4
Humidity sensor failure. [46]
8
Capacitance reference error. [47]
16
Ambient temperature out of range. [48]
32
Firmware checksum mismatch. [49]
64
Device settings corrupted. [50]
128
Additional configuration settings corrupted. [51]
256
Sensor coefficients corrupted. [52]
512
Main configuration settings corrupted. [53]
2048
Supply voltage out of range. [55]
8192
Non-volatile memory read/write failure. [57]
16384
Calibration certificate checksum mismatch. [58]
If several errors are active at the same time, the Modbus Error Code register contains the sum
of currently active error codes.
Refer to
Error Messages in Insight Software (page 70)
states.
80
Register Description
Device status
Error code
Security hash
Corresponding Error
Data
Notes
Format
16-bit
1 = No errors
integer
32-bit field
32-bit
Security hash changes on
integer
any change to device
settings/adjustments
for more information about the error