CAUTION.
Use only as specified. The probe is intended to be used only with compatible Teledyne
LeCroy instruments. Use of the probe and/or the equipment it is connected to in a
manner other than specified may impair the protection mechanisms.
Do not bend cables excessively. The input leads have a jacket wear indicator that
shows through when the jacket is excessively worn. If the white "WEAR" indicator on
the input leads is visible, cease use and contact Teledyne LeCroy service for repair or
replacement.
Use only within the operational environment listed. Do not use in wet or explosive
atmospheres.
Keep product surfaces clean and dry.
Do not operate with suspected failures. Before each use, inspect the probe and
accessories for any damage such as tears or other defects in the probe body, cable
jacket, accessories, etc. If any part is damaged, cease operation immediately and
sequester the probe from inadvertent use.
Operating Environment
Temperature, Operating
Temperature, Non-operating
Relative Humidity, Operating
Relative Humidity, Non-operating
Altitude
Usage
0 C to 50 C
- 40 C to 70 C
5% to 80% RH (Non-Condensing)
45% RH above 30 C
5% to 95% RH (Non-Condensing)
75% RH above 30 C
45% RH above 40 C
3000 m (9842 ft) max.
Derated to 2000 m (6561 ft.) when used with clip accessories
Indoors
Safety
3