Maintenance
Cleaning
Clean only the exterior surfaces of the device using a soft cloth or swab dampened with water
or 75% isopropyl alcohol solution. Do not use harsh chemicals or abrasive cleansers. Dry the
probe and accessories thoroughly before making any voltage measurements.
CAUTION. The HVD3000 probes are not waterproof. Under no circumstances
submerge the probe in liquid or allow moisture to penetrate it.
Calibration Interval
This probe has no adjustments. The recommended calibration interval is one year. A
Performance Verification Procedure is included in this manual.
Service Strategy
The HVD3000 series probes utilize fine-pitch surface mount devices. It is, therefore,
impractical to attempt repair in the field. Defective probes must be returned to a Teledyne
LeCroy service facility for diagnosis and exchange.
CAUTION. Do not remove the covers. Refer all servicing to qualified personnel.
A defective probe under warranty will be replaced with a factory refurbished probe.
Replacement probes are factory repaired, inspected, and calibrated to the same standards as a
new product.
A probe that is not under warranty can be exchanged for a factory refurbished probe for a
modest fee. You must return the defective probe in order to receive credit for the probe core.
Replacement Parts
Replacement probe accessories can be ordered through your local sales office:
Probe
HVD310x
HVD3206
HVD3605
* Individual probe accessory tips can be purchased through your regional service center. Refer to the accessory
part numbers listed earlier in this manual.
Replacement Parts
Probe Accessory Kit *
6kV Alligator Clips
6kV Alligator Clips
Part Number
PK-HV-001
PK-HVA-06
PK-HVA-06
Maintenance