Performance Verification Procedure; Required Equipment - Teledyne Lecroy HVD3000 Series Operator's Manual

High-voltage differential probes
Hide thumbs Also See for HVD3000 Series:
Table of Contents

Advertisement

Performance Verification Procedure

This procedure can be used to verify the warranted characteristics of an HVD3000 series
probe. If the product does not meet specifications, it should be returned to a Teledyne LeCroy
service center. As there are no user accessible adjustments, there is no adjustment procedure.

Required Equipment

The following table lists the test equipment and accessories (or their equivalents) that are
required for performance verification of the probe. This procedure is designed to minimize the
number of calibrated test instruments required. Because the input and output connector
types may vary on different brands and models of test instruments, additional adapters or
cables may be required. Only the parameters listed in boldface in the "Minimum requirements"
column must be calibrated to the accuracy indicated.
Description
Digital Multimeter
Function Generator
BNC Coaxial Cable
Calibration Fixture
Banana Plug Adapter
Insulated Banana Plug
Insulated Banana Couplers (2)
Minimum Requirements
DC: 0.1% accuracy
AC: 0.2% accuracy to measure
7 mVrms to 37 Vrms at 70 Hz
Output sine wave:
20 Vpp HVD310x and HVD3602
37 Vrms HVD3605 *
* May require an amplifier to
obtain required voltage, such as
Tegam 2340.
Male-to-Male 50 Ω Cable
ProBus Extender Cable
Female BNC-to-Dual Banana
Plug
Insulated BNC-to-Shrouded
Banana Plug
Performance Verification Procedure
Example Equipment
Agilent 34401A
Fluke 8842A-09
Keithley 2001
Agilent 33120A
Stanford Research DS34
Leader LAG-120B
Pomona 5697-36
Teledyne LeCroy
PROBUS-CF01
Pomona 1269
Mueller BU-00260
Mueller BU-5671-B-12-0
Mueller BU-32601-2 (R)
Mueller BU-32601-6 (B)
21

Advertisement

Table of Contents
loading

This manual is also suitable for:

Hvd3102Hvd3106Hvd3106-6m

Table of Contents