Error message
192
_CP_LCM_Manual_EN-UK final.indd 192
_CP_LCM_Manual_EN-UK final.indd 192
Meaning
The test strip is stuck or the test strip
drum does not rotate.
Press the Test button to turn the meter off.
Open the drum compartment cover and (if
applicable) pull out the test strip. Press the
red drum release button and remove the
test strip drum. Insert the test strip drum
again and close the drum compartment
cover. The meter rotates the drum and
reads the bar code. When you next turn on
the meter to perform a test, the drum
symbol and flashing bottle symbol are dis-
played. Perform a performance check. You
may use all the test strips remaining in
the drum. The 90-day use-by period will
no longer be calculated for this drum. The
drum symbol will therefore be displayed
each time you perform a test until you
insert a new test strip drum.