Accu-Chek Compact Plus Reference Manual page 183

Hide thumbs Also See for Compact Plus:
Table of Contents

Advertisement

If this happens ...
The meter is turned on,
but the display is blank.
The clock has stopped or
the clock is slow.
When the meter is turned
on, the date is shown as
0-00 and the time is
shown as 0:00.
Symbols, error messages and troubleshooting
_CP_LCM_Manual_EN-UK final.indd 183
_CP_LCM_Manual_EN-UK final.indd 183
it means:
The display is defective.
Please contact the Accu-Chek Customer
Careline free on 0800 701000 (UK) or
1800 709600 (Ireland).
The meter was exposed to a temperature
below –10 °C and the batteries are begin-
ning to freeze.
Turn the meter off. Move to a place where
the temperature is between +10 °C and
+40 °C and wait for the temperature of
the meter to adjust to this temperature.
The meter was exposed to a tempera-
ture below –10 °C and the batteries
are beginning to freeze.
Press the Test button to turn the meter
off. Move to a place where the tem-
perature is between +10 °C and
+40 °C and wait for the temperature
of the meter to adjust to this tempera-
ture.
12
183
20.6.2007 11:04:44
20.6.2007 11:04:44