If this happens ...
182
_CP_LCM_Manual_EN-UK final.indd 182
_CP_LCM_Manual_EN-UK final.indd 182
it means:
The drum compartment cover is not
properly closed.
Close the drum compartment cover. It
must close with an audible CLICK.
The ambient temperature is low and
the batteries are almost empty.
Move to a place where the tempera-
ture is between +10 °C and +40 °C
and wait for the temperature of the
meter to adjust to this temperature.
Insert batteries (new or recharged).
Condensation has affected the elec-
tronics.
Allow the meter to dry slowly.
The meter is defective.
Please contact the Accu-Chek Cus-
tomer Careline free on 0800 701000
(UK) or 1800 709600 (Ireland).
Symbols, error messages and troubleshooting