Instrument Operation; Turning The Vx500 On And Off; Vx500 Gas Reading Mode; Configuration Icons - Industrial Scientific VX500 Instruction Manual

Photoionization detector
Table of Contents

Advertisement

I
O
NSTRUMENT
PERATION
T
URNING THE
To turn the VX500 on, press and hold
membrane keypad until the instrument emits a short beep
and the VX500 welcome screen appears on the instrument
Rev1.0
display. The current revision of instrument operating
software is shown below the instrument model name.
E To Rotate 180
ENTER
Pressing
display 180 degrees to make viewing easier in applications
where the VX500 may be hung upside-down.
ISOBUTYLENE
After the welcome screen, the VX500 configuration screen
Ratio = 1.00
Cal
100ppm
will be displayed. This screen shows all parameters
Low
100ppm
currently set in the instrument including the calibration gas
High
200ppm
concentration and all current alarm level settings.
STEL
300ppm
TWA 8H 1000ppm
Following the configuration screen, the instrument will
display a brief countdown timer while the VX500
completes the automatic warm-up cycle. During this
countdown, if
9
the VX500 will enter into the configuration mode giving
you the opportunity to change the instrument's custom
settings. If no keys are pressed during warm-up, the
display will time out and begin detecting gas in the normal
operation mode.
To turn the VX500 off, press and hold
during operation until the instrument display shows
RELEASE. After releasing
will blank and all instrument operation will cease.
8
VX500 O
O
N AND
FF
on the unit's
ON/OFF
MODE
while this screen is shown will rotate the
and
are pressed simultaneously,
at any time
ON/OFF
MODE
the instrument's display
ON/OFF
MODE
VX500 G
R
M
AS
EADING
ODE
Once the VX500 enters the Gas Reading Mode the
instrument display will show the concentration of gas
currently being detected along with the corresponding gas
type. For general VOC detection this gas type will be
isobutylene. You may select another gas in the
configuration mode by editing the "favorites" list (see
page 16). Gases selected from the favorites list will use
response factors to determine actual gas concentrations.
These response factors are referenced to isobutylene
unless the VX500 is directly calibrated to the specific gas.
If the monitored concentration of gas exceeds a preset
limit for low alarm, high alarm, STEL, or TWA the
VX500 will alert you by sounding an audible tone in
conjunction with flashing an ultra bright red LED. Alarm
conditions are also activated whenever a sensor fails
calibration, the pump fails or becomes blocked, or the
battery power is too low to operate the VX500.
C
ONFIGURATION ICONS
The top line of the instrument display will show a series of
icons which identify the instrument's current operating
configuration. These icons are identified below and are
explained in greater detail in later sections of this manual.
Datalogging On
Battery Type and charge status (Shading indicates
charge status) R = rechargeable A = alkaline
Tick Mode
Snapshot Mode
R
0.1ppm
Direct Cal
ISOBUTYLENE
9

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents