Input Voltage vs. Frequency and Burn Limit

The Maximum Input Voltage curve (solid line on graph) shows the maximum voltage that can be
applied to the probe inputs without risking damage to the probe. The lower Burn Limit curve
(dashed line on graph) shows the maximum voltage that can be applied to the probe inputs
while the operator is handling the inputs. Handling the inputs while connecting the probe to
active signals above this burn limit could result in injury to the operator.
WARNING. To avoid risk of electric shock or fire, comply with the burn limits of the
derating curve when measuring high-frequency signals with hand-held accessories.
Do not exceed the voltage or category rating of the probe or accessories (whichever
is less). Keep your fingers behind the finger guard of the probe. Keep the output cable
away from the circuits being measured. Use only the specified accessories.