CALIBRATION

This scale has been factory calibrated, and does not require calibration prior to
use. If required, the scale can be calibrated. Calibration of your scale is performed
using kilograms (KG). 20 kilograms are required for calibration, applied in 5 kg and
10 kg increments.
Note: Only weights certified and traceable to national standards should be used
for calibration procedures.
With the scale off, press and hold the LB/KG button (524KL) or the
blank blue button (524KG) and ZERO buttons at the same time, then
press the ON/OFF button to turn on the scale. "CAL" will briefly
appear on the display, followed by four numbers indicating that you
can begin calibration. Release all three buttons.
Press the ZERO button. "C-5" will flash on the display, followed by
"0"or "1".
Place 5 kg on the scale. Four numbers will appear. When calibration
numbers stabilize, press the ZERO button. "C-15" will appear on the
display, followed by "0" or "1".
Place an additional 10 kg on the scale, for a total of 15 kg. Four
numbers will appear. When calibration numbers stabilize, press the
ZERO button. "C-20" will appear on the display, followed by "0" or "1".
Place an additional 5 kg on the scale, for a total of 20 kg. Four
numbers will appear. When calibration numbers stabilize, press the
ZERO button. "END" will appear on the display, followed by "OFF".
The scale will automatically shut off.
ACTION
13
DISPLAY
SHOWS
"CAL" followed by
"XXXX"
"C-5" then "0" or
"1".
"XXXX" followed
by "C-15" then "0"
or "1".
"XXXX" followed
by "C-20" then "0"
or "1".
"XXXX" followed
by "END" then
"OFF"