Meter Overview
 WARNING
This Quick Start Guide does not replace the User's
Manual for your Accu-Chek Aviva blood glucose
1
meter. The User's Manual contains important
handling instructions and additional information.
Start Here
2
Quick Start Guide
 WARNING
The lancing device is for single patient use only and
3
should not be shared.
4
1
6
1
Power/Set Button
2
Display
3
Right Arrow and Left Arrow Buttons
4
Test Strip Slot
5
Battery Door
6
Infrared (IR) Window
52459_07022425001.indd 1
Setting up the Lancing Device
1
2
3
4
Release button
Ejector
Cap
Priming
button
Penetration depth
Familiarize yourself with the
Get a lancet from the pocket in the
Remove the cap by pulling it
Insert the lancet into the lancing
Twist off the lancet's protective
5
Accu-Chek Softclix lancing
carry case.
straight off. Do not twist the cap.
device until the lancet clicks.
cover.
device.
6
7
8
9
Replace the cap on the lancing
Adjust the lancet depth to a
Press the priming button in as
Set the lancing device aside until
device. Make sure the notch on the
comfortable penetration level. Start
far as it will go, like a pen. Do not
you are ready to obtain a blood
cap lines up with the notch on the
at 2. For tougher skin, dial to a
press the release button while
sample.
lancing device.
higher number.
pressing the priming button. The
release button turns yellow when
the lancing device is ready.
Performing a Blood Glucose Test
5
1
2
3
4
Wash and dry your hands before
Check the use by date on the test
Remove a test strip from the test
Insert the test strip into the meter
you perform a blood glucose test.
strip container. Do not use test
strip container. Close the cap
in the direction of the arrows until
strips past the use by date.
tightly.
the meter beeps.
5
6
Hold the lancing device firmly
Touch the end of the test strip to
The meter beeps and flashes
The test result appears on the
against the edge of your fingertip
the blood drop. Do not put blood on
when there is enough blood in the
display after 5 seconds.
and press the yellow release
top of the test strip.
test strip.
button to poke your fingertip.
Error Messages
If any of the error messages below appear on the display at any time,
E-9 The battery is almost out of power. Change the battery now. If the
refer immediately to the User's Manual, Chapter 5, Display and Error
message reappears after the battery has been replaced, remove the
Messages.
battery again, press any meter button, then reinsert the battery.
E-1 The test strip may be damaged or not properly inserted. Remove and
E-10 The time and date settings may be incorrect. Make sure the time
reinsert the test strip, or replace it if damaged.
and date are correct and adjust, if necessary.
E-3 Your blood glucose may be extremely high or a meter or a test strip
error has occurred. If your test result matches how you feel, contact your
healthcare professional immediately. If your test result does not match
how you feel, repeat the blood glucose test and see the User's Manual,
Chapter 2, Unusual Blood Glucose Results. If the second test result does
not match how you feel, perform a control test with the control solution
A flashing blood drop symbol
and a new test strip. If the control result is within the acceptable range,
appears.
review the proper testing procedure and repeat the blood glucose test
with a new test strip. If the E-3 code still appears for your blood glucose
test, your blood glucose result may be extremely high and above the
system's reading range. Contact your healthcare professional
immediately. If the control result is not within the acceptable range, see
the User's Manual, Chapter 4, Understanding Out-of-Range Control
Results.
NOTE
Before you finish...
E-4 Not enough blood or control solution was drawn into the test strip for
Contact Accu-Chek Customer Care at 1-800-363-7949 if the
measurement or was applied after the test had started. Discard the test
error message continues.
strip and repeat the blood glucose or control test.
E-6 Blood or control solution was applied to the test strip before the
LAST UPDATE: 2013-04
flashing drop symbol appeared on the display. Discard the test strip and
repeat the blood glucose or control test.
Roche Diagnostics GmbH
E-7 An electronic error occurred, or in rare cases, a used test strip was
removed and reinserted. Turn the meter off and on, or take the battery
Sandhofer Strasse 116
out for 20 seconds and reinsert it. Perform a blood glucose or control
68305 Mannheim, Germany
test.
www.accu-chek.com
Remove the lancing device cap.
E-8 The temperature is above or below the proper range for the system.
Point the lancet away from you.
Refer to the test strip package insert for system operating conditions.
ACCU-CHEK, ACCU-CHEK AVIVA, and SOFTCLIX are
Slide out the ejector to release the
Move to an area with the appropriate conditions, wait 5 minutes, and
trademarks of Roche.
lancet.
repeat the blood glucose or control test. Do not artificially heat or cool
© 2014 Roche Diagnostics
the meter.
07022425001-0114
This file may not print or view at 100%.
Die lines and color breaks do not print.
Roche USA – 52459
V4/1 – 07022425001 – Black
Roche USA – 52459
V4/2 – 07022425001 – Cyan
Roche USA – 52459
V4/3 – 07022425001 – Magenta
Roche USA – 52459
V4/4 – 07022425001 – Yellow
1/10/14 10:53 AM

Advertisement

loading

Summary of Contents for Accu-Chek Aviva

  • Page 1 This file may not print or view at 100%. Manual for your Accu-Chek Aviva blood glucose E-1 The test strip may be damaged or not properly inserted. Remove and E-10 The time and date settings may be incorrect. Make sure the time Die lines and color breaks do not print.
  • Page 2 élevé et supérieur à l’intervalle d’analyse du système. Contactez votre REMARQUE professionnel de la santé dans les plus brefs délais. Si le résultat de Contactez Accu-Chek Soins aux Patients au 1 800 363-7949 si contrôle se situe hors de l’intervalle acceptable, reportez-vous au le message d’erreur s’affiche de nouveau.