Analog Output Overrange Behavior - Vaisala hmdw110 series User Manual

Humidity and temperature transmitters
Hide thumbs Also See for hmdw110 series:
Table of Contents

Advertisement

2 Product Overview

Analog Output Overrange Behavior

Analog outputs of the HMDW100 series transmitters have a defined behavior
when the values measured by the transmitter are outside the scaled analog
output range:
Output is clipped at the end of the scaled output range.
n
You can allow the output exceed the scaled range by 10% with the aover
n
serial command.
Output is set to error state (default 3.6 mA) if an error is active (for example,
n
due to sensor damage). You can change the error state using the aerr
command.
For configuration of the analog outputs using serial commands, see section
Analog output on page 33.