Testing and storage conditions

Temperature
Make sure that the following conditions are met so that the meter and finger pricker operate reliably and
you obtain accurate test results:
Blood glucose monitoring system without batteries, without test
cassette
Blood glucose monitoring system with batteries, without test cassette
Blood glucose monitoring system with batteries, with test cassette
H
At temperatures above +50 °C, the batteries could leak and damage the meter.
For blood glucose tests and control tests, the permitted temperature range is between +10 and +40 °C.
If the temperature is between +8 and +10 °C or between +40 and +42 °C, the meter will still allow you
to perform a test. However, the message Temperature too low or Temperature too high appears (see
Messages and problems page 122):
W
Do not use blood glucose results obtained despite this warning as a basis for making therapeutic
decisions. These test results may be incorrect. Incorrect test results can cause the wrong therapy
recommendation to be made and thus produce serious adverse health effects.
Storage
Testing and storage conditions
-25 to +70 °C
-10 to +50 °C
+2 to +30 °C
13
Temperature
117