Section VI – Customizing

Lock Option

The Lock option allows you to "lock" the BCS-G6 and prevent it from being used by
unauthorized people. When the BCS-G6 is locked, you will be required to enter an authorization
ID before each test.
To enable the Lock feature:
1. Press 3 to access Settings Menu – Menu Password (PW): 7538
2. When prompted [MENU ?], press # 9. The screen will display: [PIN? Y/N?]
3. Press the YES key to enable the Lock Pin Option, NO to cancel.
The default authorization ID number is 1234. The ID can be changed at any time.
To change the Authorization ID Number:
1. Press 9 to access Maintenance Menu – Menu Password (PW): 7538
2. When prompted [MENU ?], press # 1. - Enter (PW): 1973
The screen will display: [NEW PIN ----]
3. Enter the new 4 digit ID.
Note: The master PIN/Password is 1973 and this number can always be used to access the
menu should you forget the new PIN/Password.

Print Ohms Option

The Print Ohms option gives you the choice of whether or not to print the measured bio-
impedance value on the printouts. When you first receive the scale, the Print Ohms function is
set to OFF.
To enable the printing of the Ohms measurement:
1. Press 3 to access Settings Menu – Menu Password (PW): 7538
2. When prompted [MENU ?], press # 5. The screen will display: [OHMS MODE Y/N?]
3. Press the YES key to enable the Print Ohms Option, NO to cancel.