Running a Control Test

For control testing, you'll need your meter (with a test drum inserted), a bottle of
ACCU-CHEK Compact control solution, and a napkin or paper towel.
1. With a test drum inserted,
press
.
2. The time, date, and test strip
appear on the display. Next,
the flashing blood drop and
hand appear on the display.
Then, press SET to mark the
test as a control test.
CTRL is displayed on the
screen and stored in
memory to show that this
result is from a control test.
When marked as CTRL, the
result is not included in the
average and other trend
information (see Viewing
Trend Information ).
Chapter 2: Control Testing
Do not operate without the nose cover attached.
For more information, see Chapter 5: Maintenance
and Troubleshooting.