Chapter 2: Control Testing
Why Run Control Tests?
Control tests ensure that both the meter and the test strips are working properly and giving reliable results.
You should run a control test when:
•
The meter is new.
•
You open a new box of test strips.
•
You leave the cap off the test drum vial before
using the drum.
•
You want to check the performance of the meter
and test drum.
About the Control Solution
•
Make sure you use ACCU-CHEK Compact control
solution.*
•
When you first open the bottle, write the date on
the label – the solution is good for three months
from that date or until the expiration date on the
bottle, whichever comes first.
•
You drop the meter.
•
CTRL flashes in the meter display window.
•
Your test results don't agree with the way you feel.
•
You want to make sure you're performing tests
correctly.
•
Never use control solution that has expired.
•
Always close the control solution bottle tightly.
•
Store at 36-86°F (2-30°C). Do not freeze.
*ACCU-CHEK Compact Blue control solution
may not be available in all countries.
13