The primary display shows ---- until you press the
T button and a valid resistance reading is
obtained.
The high voltage symbol (Y) along with a primary
display of >2 V warns if voltage greater than 2 V
ac or dc is present. In this condition, the test is
inhibited. Disconnect the Tester and remove
power before proceeding.
If the Tester chirps when you press the T
button, the test is inhibited because voltage is
present at the probes.
Push and hold the T button to start the test. The
5.
t icon appears on the lower portion of the display
until you release the T button. The resistance
reading appears on the primary display until a new
test is started or a different function or range is
selected.
When resistance is higher than the maximum display
range, the Tester displays the > symbol and the
maximum resistance for the range.
.
12
Figure 6. Measuring Earth-Bond Resistance
bbv04f.eps