Infrared ear thermometer
Hide thumbs Also See for ThermoScan IRT 6030:
Table of Contents

Advertisement

3
The lens filter indicator will blink to signal one is
needed.
To achieve accurate measurements, make sure
a new, clean lens filter is in place before each
measurement.
Attach a new lens filter by pushing the
thermometer probe straight into the lens filter
inside the box and then pulling out.
Note: The Braun ThermoScan
will not work
®
unless a lens filter is attached.
4
Select age with AgeSmart
button
Press button to toggle through each age group.
noTE: You must select an age in order to take a measurement.
The thermometer is ready to take temperature
when the display looks like the images below.
6520
6030
Fit the probe snuggly into the ear canal, then
5
push and release the Start button
.
6
ExacTemp light will pulse while temperature is in progress. The light will
remain solid for 3 seconds to indicate that a successful temperature
reading has been achieved.
noTE: If the probe has been properly inserted into the ear canal
during the measurement, a long beep will sound to signal the
completed measurement.
If the probe has NOT been constantly placed in a stable position in the
ear canal, a sequence of short beeps will sound, the ExacTemp light
will go out and the display will show an error message (POS = position
error).
see « Errors and troubleshooting » section for more
information.
6
as shown below.
0-3 Mos
7
The confirmation beep indicates that an accurate temperature measurement has been
taken. The result is shown on the display.
0-3
3-36
36+
Months
0-3
3-36
36+
Months
0-3
3-36
36+
Months
age range
0-3 Months
>3-36 Months
>36 Months -
adult
>36 Mos –
>3-36 Mos
Temperature displays.
adult
9
8
For the next measurement, press Eject button
to remove and discard used lens filter, and put on a
new, clean lens filter.
NOTE: Defaults to last age setting used if you do
not change.
The Braun ThermoScan
ear thermometer turns off
®
automatically after 60 seconds of inactivity.
The thermometer can also be turned off by
pressing the Power button
.
The display will briefly flash OFF and it will
go blank.

Temperature taking hints

Always replace disposable lens filters to maintain accuracy and hygiene. The right ear
measurement may differ from the measurement taken in the left ear. Therefore, always take the
temperature in the same ear. The ear must be free from obstructions or excess earwax build-up
to take an accurate reading.
External factors may influence ear temperatures, including when an individual has:
Factor
Poor probe placement
Used lens filter
Dirty lens
green
Yellow
Red
Normal temperature
Moderate fever
High fever
≥35.8 – ≤37.4 °C
>37.4 °C
(≥96.4 - ≤99.4 °F)
(>99.4 °F)
≥35.4 – ≤37.6 °C
>38.5 °C
>37.6 – ≤38.5 °C
(≥95.7 - ≤99.6 °F)
(>101.3 °F)
(>99.6 – ≤101.3 °F)
>39.4 °C
≥35.4 – ≤37.7 °C
>37.7 – ≤39.4 °C
(>103.3 °F)
(≥95.7 – ≤99.9 °F)
(>99.9 – ≤103.0 °F)
Yes affects
7