Download Print this page

KERN CH15K20 Operating Instructions page 3

Hanging balance
Hide thumbs Also See for CH15K20:

Advertisement

8
Maintenance, upkeep, disposal
8.1
Cleaning
Only use a cloth dampened with mild suds and not aggressive
cleaning agents (solvents or similar). Please ensure that fluids
are not able to get into the device and rub off using a clean,
soft cloth. Loose sample residue/powder can be removed
carefully using a brush or hand vacuum cleaner.
8.2
Maintenance, upkeep
The operating company shall ensure that load-bearing com-
ponents are checked by a competent person at maximum
intervals of one year.
These checks can be necessary at intervals shorter than one
year depending on the conditions of use. This applies, for
example, in the case of particularly frequent use, increased
wear and corrosion.
The device may only be opened by trained service engineers
authorised by KERN.
Disconnect from the mains supply before opening.
8.3
Disposal
The operating company shall dispose of the packaging and
the device in compliance with the valid national or regional law
of the operating location.
9
Troubleshooting
The balance should be switched off for a short time following
an interruption in the programme sequence. It is then neces-
sary to repeat the weighing process from the beginning.
Interruption
Possible cause
Weight display is
The balance is not switched on.
not illuminated.
The batteries are wrongly inserted, the
batteries are empty
No batteries are attached
The weight dis-
Draught/air movement
play changes
Table/floor vibrations
continually
Vibrations on the balance suspension
Electromagnetic fields / static charging
(choose different location/switch off inter-
fering device if possible)
The weighing
The balance display is not set to zero
result is obviously
Adjustment is no longer correct.
incorrect
Great fluctuations in temperature.
Electromagnetic fields / static charging
(choose different location/switch off inter-
fering device if possible)
Switch the balance off if other error messages should appear
and then switch on again. Contact the manufacturer if the error
message does not disappear.
CH-BA-defsi-0422
7

Advertisement

loading

This manual is also suitable for:

Ch50k100Ch50k50