Device Alerts

• High Priority: These alerts require immediate operator response. The alert signal consists of a high priority sound,
which is a continuous two-beep pattern (indicated in the following table as: • • • •). Additionally, the backlights on
the buttons will provide a high priority flashing pattern consisting of a continuous, bright-to-off, two-flash pattern
(indicated in the following table as: ◊◊ ◊◊).
• Medium Priority: These alerts require prompt operator response. The alert signal consists of a medium priority
sound, which is a continuous one-beep pattern (indicated in the following table as: • •). Additionally, the backlights
on the buttons will provide a medium priority flashing pattern consisting of a continuous, bright-to-dim, one-flash
pattern (indicated in the following table as: ◊ ◊).
Alert Summary Table: The following table summarizes the alerts.
A
LERT
Service Required
Mask Alert
Auto Off
Humidifier Alert
Instant Message
12
A
V
UDIBLE
ISUAL
I
I
NDICATOR
NDICATOR
• • • •
◊◊ ◊◊
Screen displays
"Service required",
or the following
animated symbol.
• •
◊ ◊
Screen displays
"Mask alert" or
the following
animated symbol.
single beep
Screen displays
"Auto off" or the
following animated
symbol.
0
none
◊ ◊
Humidifier LED
icon will flash on
the device.
single beep
Home care
provider will
supply text to be
displayed.
D
A
P
EVICE
CTION
C
The device enters
Device failure.
the "Safe state" in
which the device
power remains on,
but the airflow is
disabled.
Alert present until
The breathing
action is taken.
circuit is
disconnected
or there is a
large air leak.
The airflow shuts
The mask has
off and the device
been removed.
enters the Standby
state approximately
45-60 seconds
after detection.
Alert present for
30 seconds or until
user acknowledges.
Only displayed when
Humidifier
both the humidifier
failure.
and therapy is on.
Only displayed when
Message from
therapy is off.
the provider.
P
OSSIBLE
ATIENT
AUSE
Press either the wheel or ramp
button to silence the alert.
Remove the power supply cord
from the device to remove power.
Plug the cord back into the
device's power inlet to restore
power. If the alert continues to
occur, contact your home care
provider.
Turn off airflow. Check your
breathing circuit connections
and reconnect the tubing if it has
come loose. Make sure your
mask is on properly before you
restart the airflow. If the alert
continues to occur, contact your
home care provider to have your
mask checked. You may need a
mask refitting.
Put your mask back on and turn
the airflow on to resume therapy.
Alert is present for 12 minutes or
until the condition is fixed. Turn
off airflow and reconnect the
humidifier to the device according
to the humidifier instructions.
If the alert continues to occur,
contact your home care provider.
Your home care provider may
send an instant message. Contact
your home care provider with any
questions.
A
CTION