WHAT DISPLAY SYMBOLS MEAN

Display
Condition/
Symbol
Cause
Appears during measure-
ment; blinks and beeps
when pulse is detected.
Battery voltage is
too low for monitor
to work properly.
Appears when systolic
and diastolic measure-
ments are within 10
mmHg of each other.
Appears if pressure
value is unstable due
to movement during
reading.
Appears when pressure
value did not increase
during cuff inflation:
1) cuff is not fastened
2) air connector is
not connected properly.
?
3
LifeSource Health Line (Toll-Free): 1-888-726-9966
Recommended
Action
Remain still and refrain
from talking; measure-
ment in progress.
Replace all batteries
with new ones.
Check for air leakage;
make sure cuff tube is
properly connected to
monitor and exhaust
rate is in increments of
2 mmHg. to 5 mmHg
Try measurement again
and remain very still.
Make sure cuff tube is
properly connected to
monitor, cuff is securely
fastened and exhaust rate
is between 2 mmHg and
5 mmHg. Make sure air
connector is inserted
properly into tube
and monitor.