Display Board Connector (X12) - GE TruSat Technical Reference Manual

Pulse oximeter
Hide thumbs Also See for TruSat:
Table of Contents

Advertisement

TruSat Technical Reference Manual

Display board connector (X12)

3–8
Pin # Signal Name
1
V_ON
2
GND
3
BACKLIGHT_CUR
4
BACKLIGHT_CUR
5
BACKLIGHT_RET_CUR
6
BACKLIGHT_RET_CUR
7
BATT+
8
GND
__________
9
LCD_EN
10
GND
11
GRN_AC_LED
12
GND
13
SPARE_KEY0
14
GND
15
SPARE_KEY1
16
GND
17
I2C_SDAT
18
GND
19
I2C_SCLK
20
GND
21
ALARM_SW
22
GND
23
ON/OFF
24
GND
25
ALARM_LED_RED
26
GND
27
ALARM_LED_YEL
28
GND
29
V_ON
30
GND
I/O
Description
PWR
+3 V regulated voltage from DC line
GND
Non-isolated ground
O
Backlight supply current
O
Backlight supply current
I
Backlight return current
I
Backlight return current
PWR
Battery voltage, isolated by ferrite
GND
Non-isolated ground
O
Enable signal for Display board regulator
GND
Non-isolated ground
O
Supply voltage for DC line power LED
GND
Non-isolated ground
I/O
Unused
GND
Non-isolated ground
I/O
Unused
GND
Non-isolated ground
2
O
I
C bus, serial data
GND
Non-isolated ground
2
O
I
C bus, serial clock
GND
Non-isolated ground
I
Alarm button signal, active high
GND
Non-isolated ground
I
Power button signal, active low
GND
Non-isolated ground
O
Alarm LED, red
GND
Non-isolated ground
O
Alarm LED, yellow
GND
Non-isolated ground
PWR
+3V regulated voltage from DC line
GND
Non-isolated ground

Hide quick links: