Attachable Mechanical Stage Type "E - Nikon LABOPHOT-POL Instructions Manual

Polarizing microscope
Hide thumbs Also See for LABOPHOT-POL:
Table of Contents

Advertisement

Iy using the clamp screw on the illuminator
"M".
Secure clamping of the tube is to be done
after completing the orientation
of analyz-
er
2)-(2).
]:: Referring to P. 7, mount the eyepiece tube
on the intermediate tube "P".
2)
Preparation
(1) Centering the lamp
CD
Turn
the half-reflecting
mirror swing-out
knob counterclockwise
to set the mirror
in the optical path.
@
Remove the L filter.
®
Place the ND filter on the stage and focus
on it using objective 1Ox.
@)
Remove
the eyepiece from
the sleeve,
looking
into the exit pupil of objective,
move the lamp housing back and forth to
form a sharp image of the lamp filament
on the diffuser of exit pupil.
®
Manipulate
the lamp centering screws to
center the filament image on the exit pupil.
®
Place the L filter.
If the image is found too dark with an
objective of 40X or higher, remove the L
filter.
(2) Orientation
of analyzer (intermediate tube
liP")
CD
Nearly focus on the ND filter on the stage
using objective 40x.
@
Set the analyzer graduation to "0".
®
Slightly
release the
intermediate
tube
clamp screw on the epi-illuminator
"M",
detect the orientation
of the intermediate
tube "P" to form the dark cross image on
the exit pupil of objective (Refer to Fig.
13), and in th is position clamp the inter-
mediate tube "P".
3) Objectives
Use the objectives CF M Plan Achromat P
series (Strain-free, 210/45).
4)
For manipulation
and microscopy, refer to
diascopic polarizing microscopy.
19
5. Attachable Mechanical, Stage: Type"
'.IE"
is
"
To attach the attachable stage on the graduated
stage, fit the tWo positioning
pins on the rear
side of the attachable stage into the two pin
holes on the graduated stage surface, and clamp
the screw using a driver or a coin.
Attachable
mechanical stage is equipped with
point counters, which are 0.2mm or 0.3mm in
pitch. These counters can be replaced by releas-
ing the head of the point counters by means of
a coin and removing the milled part of the
counters.
To make the click of the point counter free,
release the click spring nut. (Fig. 27)
~~~~\~r
7
!~.ff~~'
-Clampscrew
spring
-
I
I
'
.
nut
I
i
1"-'----
Positioning pms
Attachable
I
I
meChanica~1
.11
II
stage type E
~-="":~="'=~~.~1":==J,r=~"'~=~
,~----
-
'
---~
-}
~
Fig. 27