Process Current - Wiring Hookup; Figure 2.7 Process Current Wiring Hookup; Figure 2.8 Process Voltage Wiring Hookup A) With Sensor Excitation - Newport i8-AL Operator's Manual

I series temperature/process limit alarm controller
Table of Contents

Advertisement

2.3.4 Process Current
The figure below shows the wiring hookup for Process Current 0 – 20 mA.
When configuring your device, select Process Type in the Input Type Menu
(see Part 3).
2.3.5 Process Voltage
The figure below shows the wiring hookup for Process Voltage 0 – 100 mV,
0 – 1 V, 0 – 10 V.
Figure 2.8 a) Process Voltage
Wiring Hookup with
Sensor Excitation
RL - Voltage limited resistor, which allows to convert 24 Vdc internal excitation
voltage to the appropriate process input value. For instance: if the potentiometer
value is equal to 10 kΩ, the minimum RL is 14 kΩ for 10 V process input.
When configuring your device, select Process Type in the Input Type Menu
(see Part 3).

Figure 2.7 Process Current Wiring Hookup

(Internal and External Excitation)
b) Process Voltage Wiring
Hookup without Sensor
Excitation