Troubleshooting
The device does not turn on
• Make sure the plug is firmly fitted to the electrical outlet.
In the event of malfunction (insufficient pressure and flow) turn the unit off for 10 seconds
and turn back on.
After each cleaning, check that the unit is working properly by verifying that air still comes
out of the nozzle when the unit is turned to the "ON" position.
Unauthorized repair voids warranty.
If the device does not work properly after attempting these troubleshooting steps, contact
your authorized product distributor or Philips Respironics Customer Service at (inside the US)
1-800-962-1266, (outside the US) +33 1 47 52 30 00.
Maintenance and repairs
Never open the device. There are no serviceable parts within the unit. The compressor requires
no lubrication or maintenance.
Technical specifications
Electricity supply
120V/60Hz 1.7 Amp
Thermal overload protector
Thermal fuse; functioning temperature 150 C°
Compressor max. pressure
303 kPa (44 psi)
Average flow rate
7 LPM @ 10 psig
Max flow rate
9.4 LPM
Weight
3.3 lbs
Size
4.2 (H) x 6.5 (W) x 6.5 (L) inches
Noise level
58 ±3dBA
Class II device, Type B device (device with specific protection against electrical hazards).
DOUBLE INSULATED
Reference to standards
Electric safety standards IEC 60601-1
Electromagnetic compatibility according to IEC 60601-1-2
Ambient conditions
Storage conditions
Temperature
MIN -25 °C – MAX 70 °C (-13 °F – 158 °F)
Humidity
MIN 10% RH – MAX 95% RH
Operating conditions
Temperature
MIN 5 °C – MAX 40 °C (45 °F –104 °F)
Humidity
MIN 10% RH – MAX 95% RH
Altitude
From 0 – 2000 m above sea level
Operating atmospheric pressure
Performance may vary depending on altitude above sea level,
barometric pressure and temperature
Symbol glossary
l
SN
ON (power)
~
O
OFF (power)
ATTENTION
Follow instructions for use
Warranty
Respironics New Jersey, Inc. warrants the compressor to be free from defects in materials and workmanship
under normal use and operation for a period of 5 years from date of purchase from Respironics New Jersey,
Inc. The warranty is limited to repair or replacement, at Respironics New Jersey, Inc. ' s sole option, of any such
component or equipment claimed to be defective when claim is shown to be bona fide by evaluation by
Respironics New Jersey, Inc. This warranty does not extend to any components or equipment subjected to
misuse, improper operation, accidental damage, or unauthorized repairs, and is not extended to charges of, or
for, labor repairs. All items returned must be properly packaged and shipped, prepaid, by the product distributor
servicing the unit. Respironics New Jersey, Inc. shall not be liable to purchaser or others for loss of use of
equipment or for indirect, or incidental or consequential damages that might arise.
ENGLISH
Serial number
Class II, double insulated
Alternating current
Type B applied parts
Indoor use only
Manufactured by