3-1 Controller
3-1-1-5 Setting a Dip Switch
♦ SW3
SW3-1: Input function change
SW3-1
ON
Functions as trigger input.
OFF
Functions as reset input.
SW3-2: Operation mode
SW3-2
ON
Starts in the Self-Execution mode when the power supply turns ON.
OFF
Starts in the Command Execution mode when the power supply turns ON.
SW3-3: Verification at writing
SW3-3
ON
Enables verification.
OFF
Disables verification.
SW3-4: Protocol designation
SW3-4
ON
Uses the 1:N protocol.
OFF
Uses the 1:1 protocol.
Set SW3-5, SW3-6, SW3-7 and SW3-8 to OFF always. Otherwise, any operation cannot be guaranteed.
3-4