7. Reposition battery terminal covers over battery post(s).
8. Using new tie‐wraps, secure the terminal covers to the battery terminals as shown in 
FIGURE 9.3.
9. Repeat STEPS 1‐8 to install and connect the rear battery to the rear battery harness.
NEGATIVE (-)
Battery Cable
NEGATIVE (-)
Battery Terminal
FIGURE 9.3 Connecting/Disconnecting the Battery Wiring Harness

Disconnecting

NOTE: For this procedure, refer to FIGURE 9.3 on page 61.
1. Remove the existing tie‐wraps that secure the battery terminal covers to the battery 
terminals.
2. Peel back RED battery terminal cover to expose RED battery cable connection to 
battery terminal.
3. Peel back BLACK battery terminal cover from BLACK battery cable on front battery 
or GREY battery terminal cover from BLACK battery cable on rear battery.
4. Remove the mounting screws and nuts that secure the POSITIVE (+) RED battery cable 
to the POSITIVE (+) battery terminal/post as shown in FIGURE 9.3.
5. Remove the mounting screws and nuts that secure the NEGATIVE (‐) BLACK battery 
cable to the NEGATIVE (‐) battery terminal/post as shown in FIGURE 9.3.
6. Set wiring harness aside.
7. Repeat STEPS 1‐6 to disconnect the rear battery from the rear battery harness.
Part No 1143153
Nut
Mounting
Screw
Tie-Wrap
Battery
61
Pronto® M91™Base with SureStep®
SECTION 9—BATTERIES
POSITIVE (+)
Battery Cable,
Mounting Screw with
Fuse Hardware
Tie-Wrap
Locknut
POSITIVE (+)
Battery Terminal

Hide quick links: