Push Buttons - Special Functions; Reset To "Normal" Display; Display Only; Default Values - GE DewPro MMY31 Installation And Operation Manual

General eastern trace moisture transmitter
Hide thumbs Also See for DewPro MMY31:
Table of Contents

Advertisement

September 2005
Push Buttons - Special
Functions

Reset to "Normal" Display

Display Only

Default Values

Changing Values

4-4
Pressing the V and H buttons simultaneously returns the user to
VH 00 (normal display).
Five (5) matrix locations are for display only and may not be changed
by the user (see Table 4-1 on page 4-3). The 'display only' fields are
as follows:
VH 00
Normal display (in dewpoint or ppm
VH 90
During a system alarm, displays the error code for the
fault encountered.
VH 91
During normal operation, displays the previous error
code for reference.
VH 92
Displays the factory issued identification number.
VH 93
Displays the factory-issued reference number
designating the device type and software version.
A default value is assigned to each programmable matrix field. The
values are present after a reset to factory programmed data has been
executed (see VH 95).
When unlocked (VH89 = 50), values in certain matrix locations can
be changed using the + and – buttons. The changeable digit flashes.
).
v