6. Error Displays

Error Display
Cause
Your feet were not positioned over
the electrodes.
The measurement position was not
stable or your feet were not placed
correctly.
Feet are too dry.
The body fat percentage and BMI
values are outside the
measurement range noted in the
specifications on page 23.
Your feet were not positioned over
the electrodes.
An operating error has occurred.
You stepped onto the measurement
platform before 0.0 kg was
displayed.
The unit was moved before 0.0 kg
was displayed.
Movement during measurement.
Your weight is over 150.0 kg.
Correction
Make sure that you are standing on
the measurement platform correctly
and try again.
Remain still and do not move during
measurement.
Slightly moisten the soles of your
feet with a damp towel and try
again.
Check that the registered settings
for height, age and gender are
correct.
If this error still appears and the
height, age, and gender settings
are correct, measurements cannot
be taken.
Make sure that you are standing on
the measurement platform correctly
and try again.
Remove the batteries then reinsert
them after waiting for at least one
minute.
Push the power switch to turn the
unit on, and repeat the
measurement.
If this error still occurs, consult your
OMRON service representative.
Wait until 0.0 kg is displayed before
stepping onto the measurement
platform.
Do not move the unit until 0.0 kg is
displayed.
Do not move till measurement is
complete.
This unit can only be used by
people weighing less than 150.0 kg.
6.Error Displays
EN
21