Chapter 3 Pre Installation; Installation Requirements; Feedwater Requirements; Feedwater Connection Needed - Millipore DIRECT-Q 3 SYSTEM User Manual

Water purification system
Table of Contents

Advertisement

Chapter 3 P
3-1 I
NSTALLATION
F
R
EEDWATER
EQUIREMENTS
Type of Feedwater
Flowrate
Minimum Pressure
Maximum Pressure
Conductivity
Temperature
Fouling Index
Aluminum
Manganese
Free Chlorine
Langelier Saturation Index
F
C
EEDWATER
ONNECTION
Feedwater Piping Connection
R
F
R
EJECT
LOW
EQUIREMENT
Drain Capacity
I
RE
NSTALLATION
R
EQUIREMENTS
pH
Iron
TOC
N
EEDED
P
Potable
≥ 40 LPH
≥ 0.5 bar
≤ 6 bar
< 2000 μS/cm
5 to 35 °C
4 - 10
< 10
< 0.1 ppm as CaCO
< 0.05 ppm as CaCO
< 0.05 ppm as CaCO
< 1 ppm
< +0.2
< 2000 ppb
1/2 inch Male GAZ, NPTM or BSPM
30 LPH
- 6 -