Device Alerts

• High Priority: These alerts require immediate operator response. The alert signal consists of a high priority sound,
which is a continuous two-beep pattern (indicated in the following table as: • • • •). Additionally, the backlights on
the buttons will provide a high priority flashing pattern consisting of a continuous, bright-to-off, two-flash pattern
(indicated in the following table as: ◊◊ ◊◊).
• Medium Priority: These alerts require prompt operator response. The alert signal consists of a medium priority
sound, which is a continuous one-beep pattern (indicated in the following table as: • •). Additionally, the backlights
on the buttons will provide a medium priority flashing pattern consisting of a continuous, bright-to-dim, one-flash
pattern (indicated in the following table as: ◊ ◊).
alert Summary Table: The following table summarizes the alerts.
A
lert
i
ndicAtor
Service Required
Mask Alert
Auto Off
single beep
Humidifier Alert
12
A
V
udible
isuAl
i
ndicAtor
• • • •
◊◊ ◊◊
Screen displays
"Service required",
indicating that
service is required.
• •
◊ ◊
Screen displays
"Mask alert".
Screen displays
"Auto off".
◊ ◊
none
Humidifier LED
icon will flash.
d
A
P
eVice
ction
ossible
c
The device enters
Device failure.
the "Safe state" in
which the device
power remains on,
but the airflow is
disabled.
Alert present until
The breathing
action is taken.
circuit is
disconnected
or there is a
large air leak.
The airflow shuts
The mask has
off and the device
been removed.
enters the Standby
state approximately
45-60 seconds
after detection.
Alert present for
30 seconds or until
user acknowledges.
Only displayed when
Humidifier
both the humidifier
failure.
and therapy is on.
P
Atient
Ause
Press either the wheel or ramp
button to silence the alert.
Remove the power supply cord
from the device to remove power.
Plug the cord back into the
device's power inlet to restore
power. If the alert continues to
occur, contact your home care
provider.
Turn off airflow. Check your
breathing circuit connections
and reconnect the tubing if it has
come loose. Make sure your
mask is on properly before you
restart the airflow. If the alert
continues to occur, contact your
home care provider to have your
mask checked. You may need a
mask refitting.
Put your mask back on and turn
the airflow on to resume therapy.
Alert is present for 12 minutes or
until the condition is fixed. Turn
off airflow and reconnect the
humidifier to the device according
to the humidifier instructions.
If the alert continues to occur,
contact your home care provider.
A
ction
User Manual

Advertisement

Table of Contents

   Also See for Philips REMstar Auto A-Flex

This manual is also suitable for:

Respironics remstar auto a-flex

Comments to this Manuals

Symbols: 0
Latest comments:
  • Robert Dec 14, 2016 01:18:
    Good product it will health care in the world
  • Ann Jarvella Wilson Mar 01, 2015 06:40:
    If I cannot attach an extension to my new REMstar Auto A-Flex machine, how can I use it when not near an outlet? for example, in a hotel room or when visiting in homes that don't have appropriat​ely located outlets?

    ​Can I use a power strip? or a heavy duty extension cord? or ???

    Can anyone give suggestion​s?
    Thanks!