  • Page 1 BiPAP Pro Bi-Flex USER MANUAL...
  • Page 2 © 2012 Koninklijke Philips Electronics N.V. All rights reserved.
  • Page 3: Table Of Contents

    Traveling with the System ..........................20 Cleaning the Device ............................20 Cleaning or Replacing the Filters ........................20 Cleaning the Tubing ..............................20 Service ..................................20 Specifications ................................21 Disposal ..................................22 How to Contact Philips Respironics .......................22 EMC Information ..............................23 Limited Warranty ........................... Back Page User Manual...
  • Page 4: Intended Use

    Several accessories are available to make your OSA treatment with the BiPAP Pro Bi-Flex system as convenient and comfortable as possible. To ensure that you receive the safe, effective therapy prescribed for you, use only Philips Respironics accessories.
  • Page 5: Cautions

    • When DC power is obtained from a vehicle battery, the device should not be used while the vehicle’s engine is running. Damage to the device may occur. • Only use a Philips Respironics DC Power Cord and Battery Adapter Cable. Use of any other system may cause damage to the device. Contraindications...
  • Page 6: Symbol Key

    Symbol Key The following symbols may appear on the device and power supply: ymbol efinition Consult accompanying instructions for use. AC Power DC Power IP22 Drip Proof Equipment Caution, consult accompanying documents. ESD Warning symbol Class II (Double Insulated) Type BF Applied Part For Indoor Use Only.
  • Page 8: System Contents

    System Contents Your BiPAP Pro Bi-Flex system may include the following items: • Device • SD card • User manual • Side cover panel (optional) • Carrying case • Reusable gray foam filter • Flexible tubing • Disposable ultra-fine filter (optional) •...
  • Page 9: Control Buttons

    • CPAP – This mode delivers Continuous Positive Airway Pressure; CPAP maintains a constant level of pressure throughout the breathing cycle. Available Flex Comfort Features The BiPAP Pro Bi-Flex device offers the following optional Flex comfort features: • Bi-Flex – Provides pressure relief upon exhalation to improve comfort based on your needs during Bi-level therapy.
  • Page 10: Installing The Air Filters

    To connect your breathing circuit to the device, complete the following steps: 1. Connect the flexible tubing to the air outlet on the side of the device. Note: Make sure the Tubing type setting (15 or 22) matches the tubing you are using (Philips Respironics 15 or 22 mm tubing).
  • Page 11: Navigating The Device Screens

    Complete the following steps to operate the device using AC power: 1. Plug the socket end of the AC power cord (included) into the power supply (also included). IMPORTANT: When you are using Heated Tubing with the compatible System One Heated Humidifier, you must use the 80W power supply.
  • Page 12: Ramp Feature

    5. Make sure that no air is leaking from your mask into your eyes. If necessary, adjust the mask and headgear until the air leak stops. See the instructions provided with your mask for more information. Note: A small amount of mask leak is normal and acceptable. Correct large mask leaks or eye irritation from an air leak as soon as possible.
  • Page 13: Flex/Rise Time Screen

    Flex/Rise time Screen From the Home screen, highlight “Flex” or “Rise time” and press the wheel. The following screen will appear. Note: This screen will be blank on the Home screen if your provider has not enabled Flex or Rise time on your device. Flex Screen Note: “Bi-Flex”...
  • Page 14: Setup Screen

    Setup Screen From the Home screen, highlight “Setup” and press the wheel. The following Setup screen will appear. The user can change settings in the Setup menu. Back Flex 1 2 3 Rise time 1 2 3 Heated Tube humidification on off SYSTEM ONE humidification on off Humidifier...
  • Page 15 • Tubing Type - This setting allows you to select the correct size diameter tubing that you are using with the device. You can choose either (22) for the Philips Respironics 22 mm tubing, or (15) for the Philips Respironics 15 mm tubing.
  • Page 16: Info Screen

    Info Screen From the Home screen, highlight “Info” and press the wheel. The following Info screen will appear. The user cannot change settings in the Info menu. Note: These screens are only for reference. Your home care provider may periodically ask you for this information. Back Status Phone-in...
  • Page 17 • AHI - The device accumulates individual Apnea/Hypopnea indices (AHI) for each session the patient used the device. This screen displays the nightly AHI value for the most recent 1 day time frame. It also displays the average of these individual nightly AHI values over a 7 day and a 30 day time frame (provided the device has at least 7 or 30 days of data respectively).
  • Page 18: Device Alerts

    Device Alerts • High Priority: These alerts require immediate operator response. The alert signal consists of a high priority sound, which is a continuous two-beep pattern (indicated in the following table as: • • • •). Additionally, the backlights on the buttons will provide a high priority flashing pattern consisting of a continuous, bright-to-off, two-flash pattern (indicated in the following table as: ◊◊...
  • Page 19 lert udible isuAl eVice ction ossible Atient ction ndicAtor Ause ndicAtor Heated Tube Error none Humidifier LED Alert present for Tubing may be Alert is present for 30 seconds or icon will slowly 30 seconds or until overheating or until the condition is fixed. Turn off flash orange for condition is fixed.
  • Page 20 lert udible isuAl eVice ction ossible Atient ction ndicAtor Ause ndicAtor • • ◊ ◊ SD Card: Corrupt Alert present until A problem Choose “yes” to reformat action is taken. exists with the the card. Screen displays Screen displays SD card. The “Reformatting...
  • Page 21: Troubleshooting

    Make sure the Tubing type setting (22 or 15) matches the tubing that you are using feels too high or too setting may be (Philips Respironics 22 or 15 mm tubing). low. incorrect. If you are using the Heated Tubing, this setting will be 15H and you cannot change it.
  • Page 22: Accessories

    Using the SD Card The BiPAP Pro Bi-Flex system comes with an SD card inserted in the SD card slot on the back of the device to record information for the home care provider. Your home care provider may ask you to periodically remove the SD card and send it to them for evaluation.
  • Page 23: Traveling With The System

    Contact your home care provider for additional information. Airline Travel The BiPAP Pro Bi-Flex device is suitable for use on airlines when the device is operating from an AC or DC power source.
  • Page 24: Specifications

    Specifications Environmental Operating Temperature: 5° to 35° C (41° to 95° F) Storage Temperature: -20° to 60° C (-4° to 140° F) Relative Humidity (operating & storage): 15 to 95% (non-condensing) Atmospheric Pressure: 101 to 77 kPa (0 - 2286 m / 0 - 7500 ft) Physical Dimensions: 18 x 14 x 10 cm (7”...
  • Page 25: Disposal

    How to Contact Philips Respironics To have your device serviced, contact your home care provider. If you need to contact Philips Respironics directly, call the Philips Respironics Customer Service department at 1-800-345-6443 or 1-724-387-4000. You can also use the following address: Respironics, Inc.
  • Page 26: Emc Information

    EMC Information Guidance and Manufacturer’s Declaration - Electromagnetic Emissions – This device is intended for use in the electromagnetic environment specified below. The user of this device should make sure it is used in such an environment. miSSionS ompliance lectromagnetic nvironment uiDance RF emissions...
  • Page 27 Guidance and Manufacturer’s Declaration - Electromagnetic Immunity – This device is intended for use in the electromagnetic environment specified below. The user of this device should make sure it is used in such an environment. iec 60601 t mmunity evel ompliance evel lectromagnetic...
  • Page 28: Limited Warranty

    Limited Warranty Respironics, Inc. warrants that the system shall be free from defects of workmanship and materials and will perform in accordance with the product specifications for a period of two (2) years from the date of sale by Respironics, Inc. to the dealer.